Performance of the automated DNA extraction with MagNA Pure 24 for further genetic testing for hemoglobinopathies with Globin StripAssays

Georg Leixner, Astrid Voill-Glaninger, Julia Gaugeler-Kurzweil, Tanja Kusstatscher, Tanja Panic-Jankovic, Wolfgang Novak, Leo Kager, Christian Oberkanins, André Viveiros

Wiener klinische Wochenschrift · 2025

DOI: 10.1007/s00508-025-02674-9 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Wiener klinische Wochenschrift, 2025. Verfasst von Georg Leixner und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Laboratory Medicine, Klinik Landstrasse, Vienna Healthcare Group, Juchgasse 25, 1030, Vienna, Austria. Hauptthemen laut MEDLINE: Genetic Testing, Hemoglobinopathies, DNA, beta-Globins. Die Arbeit nennt 22 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: Diagnosing hemoglobinopathies through reverse hybridization with Globin StripAssays (ViennaLab, Vienna, Austria) is widely performed. There are only two authorized hands-on DNA extraction methods for these assays: the spin Micro DNA Extraction Kit (ViennaLab, Vienna, Austria) for the α‑Globin StripAssay and GENXTRACT Blood DNA Extraction System (ViennaLab, Vienna, Austria) for the β‑Globin StripAssay. We set out to evaluate the performance of the automated DNA extraction with MagNA Pure 24 (Roche, Rotkreuz, Switzerland) for further testing with Globin StripAssays. METHODS: We analyzed blood samples of 41 patients with clinical suspicion of hemoglobinopathies. As the gold standard we stipulated the ViennaLab DNA extraction methods and compared them to the MagNA Pure 24 Total Isolation Kit performed on MagNA Pure. The results of the different test strips were qualitatively classified with respect to good and weak read outs, background signals as well as true/false positives or negatives. RESULTS: A total of 6226 StripAssay lines were analyzed. After extraction with the gold standard 98.3% had a good read out, 0.9% a weak read out and in 0.9% a background signal was identified. After extraction with MagNA Pure 99.4% had a good read out, 0.2% a weak read out and a background signal was found in only 0.5% of the lines (P < 0.001) and there were no false positives or negatives (sensitivity and specificity 100.0%). CONCLUSION: The automated DNA extraction with MagNa Pure showed noninferior results when compared with the Spin Micro and GENXTRACT DNA extraction kits enabling the usage of the MagNa Pure for the purpose of DNA extraction for further genetic testing with Globin StripAssays.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Genetic TestingHemoglobinopathiesDNAbeta-GlobinsHumansSensitivity and Specificitybeta-ThalassemiaReagent Kits, Diagnosticalpha-Globins

Stichwörter der Autor:innen

Alpha-thalassemiaBeta-thalassemiaHemoglobin genesHemoglobin variantsViennaLab

Substanzen

DNAReagent Kits, Diagnosticbeta-Globinsalpha-Globins

Autor:innen und Institutionen

  • Institute for Laboratory Medicine, Klinik Landstrasse, Vienna Healthcare Group, Juchgasse 25, 1030, Vienna, Austria.
  • Department of Pediatrics and Medical University Vienna Children's Cancer Research Institute, St. Anna Children's Hospital, Vienna, Austria.
  • St. Anna Children's Cancer Research Institute, Vienna, Austria.
  • ViennaLab Diagnostics GmbH, Gauzendorfer Gürtel 43-45, 1120, Vienna, Austria.
  • Institute for Laboratory Medicine, Klinik Landstrasse, Vienna Healthcare Group, Juchgasse 25, 1030, Vienna, Austria. andre.viveiros@gesundheitsverbund.at.

Zitieren

APA

Leixner, G., Voill-Glaninger, A., Gaugeler-Kurzweil, J., Kusstatscher, T., Panic-Jankovic, T., Novak, W., Kager, L., Oberkanins, C., & Viveiros, A. (2025). Performance of the automated DNA extraction with MagNA Pure 24 for further genetic testing for hemoglobinopathies with Globin StripAssays. Wiener klinische Wochenschrift. https://doi.org/10.1007/s00508-025-02674-9

Vancouver

Leixner G, Voill-Glaninger A, Gaugeler-Kurzweil J, Kusstatscher T, Panic-Jankovic T, Novak W, et al. Performance of the automated DNA extraction with MagNA Pure 24 for further genetic testing for hemoglobinopathies with Globin StripAssays. Wiener klinische Wochenschrift. 2025. doi:10.1007/s00508-025-02674-9

BibTeX
@article{Leixner2025,
  author = {Leixner, Georg and Voill-Glaninger, Astrid and Gaugeler-Kurzweil, Julia and Kusstatscher, Tanja and Panic-Jankovic, Tanja and Novak, Wolfgang and Kager, Leo and Oberkanins, Christian and Viveiros, André},
  title = {Performance of the automated DNA extraction with MagNA Pure 24 for further genetic testing for hemoglobinopathies with Globin StripAssays},
  journal = {Wiener klinische Wochenschrift},
  year = {2025},
  doi = {10.1007/s00508-025-02674-9}
}
Kostenlos starten →