Evaluation of the envision endoscopy SimpleStitch suturing system for closure of gastrointestinal defects in a porcine model
Surgical Endoscopy · 2026-04 · 40 (4) · 3510-3518
DOI: 10.1007/s00464-026-12718-4 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Surgical Endoscopy, online erschienen am 10.03.2026. Verfasst von Manik Aggarwal und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, MN, USA. Hauptthemen laut MEDLINE: Suture Techniques, Endoscopy, Gastrointestinal, Stomach. Die Arbeit nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wake Forest University.
Abstract (Originalsprache)
Background
Flexible endoscopic suturing tools are complex and may have a long learning curve. This porcine study evaluated the safety and performance of a simplified suturing system compared with a commercially available device for the repair of gastrointestinal mucosal defects.
Methods
This IACUC-approved study included four healthy swine. A total of ten defects (six in the stomach and four in the rectosigmoid colon) were created in each animal. Defects were randomly assigned to closure with either the novel or the commercially available system using a therapeutic gastroscope. Technical success was defined as mucosal closure of the defect with the inability to visualize any significant portion of the resection bed. Additional performance metrics included procedure time and device ease of use (assessed using the NASA Task Load Index [TLI]) and adverse events.
Results
No adverse events were reported post-procedurally for any of the test animals. The proportion of target resection sites achieving technical success was 100% in both treatment groups. The mean SimpleStitch NASA-TLI score was lower compared to the OverStitch device. Closure times were similar between the two devices. Histological assessment scores indicated expected healing response without evidence of perforation, leakage, or abscess formation.
Conclusion
A novel full-thickness suturing system safely and effectively closed mucosal defects. Lower NASA-TLI scores suggest that the novel suturing device may offer simpler, less demanding use compared to the predicate device, potentially reducing the learning curve for endoscopic suturing procedures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, MN, USA.
- Developmental Endoscopy Unit, Mayo Clinic, Rochester, MN, USA.
- Developmental Endoscopy Unit, Mayo Clinic, Rochester, MN, USA. astorm@wakehealth.edu.
- Division of Gastroenterology, Wake Forest University School of Medicine, Winston-Salem, NC, USA. astorm@wakehealth.edu.
Förderung
- Wake Forest University
Zitieren
APA
Aggarwal, M., AbiMansour, J. P., Matar, R., Rugivarodom, M., Salameh, Y., Zeid, H. A., Kamba, S., Gades, N. M., Rajan, E., & Storm, A. C. (2026). Evaluation of the envision endoscopy SimpleStitch suturing system for closure of gastrointestinal defects in a porcine model. Surgical Endoscopy, 40(4), 3510-3518. https://doi.org/10.1007/s00464-026-12718-4
Vancouver
Aggarwal M, AbiMansour JP, Matar R, Rugivarodom M, Salameh Y, Zeid HA, et al. Evaluation of the envision endoscopy SimpleStitch suturing system for closure of gastrointestinal defects in a porcine model. Surgical Endoscopy. 2026;40(4):3510-3518. doi:10.1007/s00464-026-12718-4
BibTeX
@article{Aggarwal2026,
author = {Aggarwal, Manik and AbiMansour, Jad P and Matar, Reem and Rugivarodom, Manus and Salameh, Yara and Zeid, Hadi Abou and Kamba, Shunsuke and Gades, Naomi M and Rajan, Elizabeth and Storm, Andrew C},
title = {Evaluation of the envision endoscopy SimpleStitch suturing system for closure of gastrointestinal defects in a porcine model},
journal = {Surgical Endoscopy},
year = {2026},
volume = {40},
number = {4},
pages = {3510--3518},
doi = {10.1007/s00464-026-12718-4}
}
EUVIMED