EUVIMEDEUVIMED entdecken
Multizentrische StudieVergleichsstudie

Comparison of outcomes after transcatheter patent ductus arteriosus closure in infants weighing ≤ 1.5 kg and > 1.5–4 kg

Katia Librandi, Carles Bautista-Rodriguez, Gianfranco Butera, Giuseppe Annoni, Paul Padovani, Sandrine Foldvari, Alban-Elouen Baruteau, Alain Fraisse

European Journal of Pediatrics · 29.10.2026 · 185 (11)

DOI: 10.1007/s00431-026-07581-3 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift European Journal of Pediatrics, online erschienen am 29.10.2026. Verfasst von Katia Librandi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Pediatric Cardiology, Royal Brompton & Harefield NHS Foundation Trust, Sydney Street, London, SW3 6NP, UK. Hauptthemen laut MEDLINE: Cardiac Catheterization, Ductus Arteriosus, Patent, Septal Occluder Device. Die Arbeit nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Transcatheter closure of patent ductus arteriosus (PDA) with the Amplatzer Piccolo Occluder is safe and effective in infants ≤ 2 kg. In larger infants, off-label devices are often required, but outcome data are limited. We evaluated outcomes in infants ≤ 4 kg, comparing the ≤ 1.5 and > 1.5–4 kg groups. This retrospective, international, multicenter study included 358 infants who underwent transcatheter PDA closure at 5 tertiary centers between 2016 and 2024: 191 weighed ≤ 1.5 kg (group 1) and 167 > 1.5–4 kg (group 2). Median age was 39 days (interquartile range 25–61) and median weight was 1.48 (1.1–2.2) kg; 93.6% were preterm. Procedural success was higher in group 1 (98.4% vs 94.0%; p = 0.044). Major adverse events were more frequent in group 2 (9.6% vs 3.1%; p = 0.0144; odds ratio = 3.27; 95% confidence interval, 1.25–8.55). Device embolization occurred only in group 2 (5.4% vs 0%; p 1.5–4 kg group has a higher risk of major adverse events, particularly device embolization. Better patient selection and dedicated devices may improve safety. What is Known: • Transcatheter PDA closure with the Amplatzer Piccolo Occluder is safe and effective in very small infants . • In larger infants, alternative devices are often used off-label, but evidence on procedural outcomes remains limited . What is New: • Infants weighing >1.5–4 kg had lower procedural success and more major adverse events than those weighing ≤1.5 kg . • Device embolization occurred exclusively in the >1.5–4 kg group, supporting careful anatomical assessment and the development of dedicated devices .

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac CatheterizationDuctus Arteriosus, PatentSeptal Occluder DeviceFemaleHumansInfantInfant, NewbornMaleRetrospective StudiesTreatment Outcome

Stichwörter der Autor:innen

Device embolizationInterventional congenital cardiologyPatent ductus arteriosusProcedural outcomesTranscatheter PDA closure

Autor:innen und Institutionen

  • Pediatric Cardiology, Royal Brompton & Harefield NHS Foundation Trust, Sydney Street, London, SW3 6NP, UK.
  • Cardiology, Cardiac Surgery and Heart Lung Transplantation, Bambino Gesù Hospital and Research Institute, IRCCS, Rome, Italy.
  • Pediatric Cardiology, Regina Margherita Children Hospital, Piazza Polonia 96, Turin, Italy.
  • Pediatric Cardiology, Grande Ospedale Metropolitano Niguarda, Milan, Italy.
  • Department of Pediatric Cardiology and Pediatric Cardiac Surgery, FHU PRECICARE, Nantes Université, CHU Nantes, 44000, Nantes, France.
  • Department of Pediatric Cardiology and Pediatric Cardiac Surgery, CIC 1413, INSERM, Nantes Université, CHU Nantes, 44000, Nantes, France.

Zitieren

APA

Librandi, K., Bautista-Rodriguez, C., Butera, G., Annoni, G., Padovani, P., Foldvari, S., Baruteau, A. E., & Fraisse, A. (2026). Comparison of outcomes after transcatheter patent ductus arteriosus closure in infants weighing ≤ 1.5 kg and > 1.5–4 kg. European Journal of Pediatrics, 185(11). https://doi.org/10.1007/s00431-026-07581-3

Vancouver

Librandi K, Bautista-Rodriguez C, Butera G, Annoni G, Padovani P, Foldvari S, et al. Comparison of outcomes after transcatheter patent ductus arteriosus closure in infants weighing ≤ 1.5 kg and > 1.5–4 kg. European Journal of Pediatrics. 2026;185(11). doi:10.1007/s00431-026-07581-3

BibTeX
@article{Librandi2026,
  author = {Librandi, Katia and Bautista-Rodriguez, Carles and Butera, Gianfranco and Annoni, Giuseppe and Padovani, Paul and Foldvari, Sandrine and Baruteau, Alban-Elouen and Fraisse, Alain},
  title = {Comparison of outcomes after transcatheter patent ductus arteriosus closure in infants weighing ≤ 1.5 kg and > 1.5–4 kg},
  journal = {European Journal of Pediatrics},
  year = {2026},
  volume = {185},
  number = {11},
  doi = {10.1007/s00431-026-07581-3}
}