Validation of smartwatch electrocardiogram intervals in children compared to standard 12 lead electrocardiograms

Julia Ernstsson, Birgitta Svensson, Petru Liuba, Constance G Weismann

European Journal of Pediatrics · 2024

DOI: 10.1007/s00431-024-05648-7 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift European Journal of Pediatrics, 2024. Verfasst von Julia Ernstsson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Sciences Lund, Lund University, Lund, Sweden. Hauptthemen laut MEDLINE: Arrhythmias, Cardiac, Electrocardiography. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lund University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Lay people are now able to obtain one-lead electrocardiograms (ECG) using smartwatches, which facilitates documentation of arrhythmias. The accuracy of smartwatch derived ECG intervals has not been validated in children though. Home-based monitoring of ECG intervals using a smartwatch could improve monitoring of children, e.g. when taking QTc prolonging medications. The aim of this study was to validate the ECG intervals measured by smartwatch in comparison to standard 12-lead ECGs in children and adolescents. Prospective study of children (age 5—17 years) at the outpatient clinic of a national pediatric heart center. Patients underwent a smartwatch ECG (ScanWatch, Withings) and a simultaneous standard 12-lead ECG. ECG intervals were measured both automatically and manually from the smartwatch ECG and the 12-lead ECG. Intraclass correlation coefficients and Bland–Altman plots were performed. 100 patients (54% male, median age 12.9 (IQR 8.7–15.6) were enrolled. The ICC calculated from the automated smartwatch and automated 12-lead ECG were excellent for heart rate (ICC 0.97, p < 0.001), good for the PR and QT intervals (ICC 0.86 and 0.8, p < 0.001), and moderate for the QRS duration and QTc interval (ICC 0.7 and 0.53, p < 0.001). When using manual measurements for the smartwatch ECG, validity was improved for the PR interval (ICC 0.93, p < 0.001), QRS duration (ICC 0.92, p < 0.001), QT (ICC 0.95, p < 0.001) and QTc interval (ICC 0.84, p < 0.001). Conclusion : Automated smartwatch intervals are most reliable measuring the heart rate. The automated smartwatch QTc intervals are less reliable, but this may be improved by manual measurements. What is Known: In adults, smartwatch derived ECG intervals measured manually have previously been shown to be accurate, though agreement for automated QTc may be fair. What is New: In children, automated smartwatch QTc intervals are less reliable than RR, PR, QRS and uncorrected QT interval. Accuracy of the QTc can be improved by peroforming manual measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arrhythmias, CardiacElectrocardiographyHumansChildMaleFemaleProspective StudiesAdolescentChild, PreschoolHeart RateReproducibility of ResultsElectrocardiography, AmbulatoryWearable Electronic Devices

Stichwörter der Autor:innen

Congenital heart diseaseElectrocardiogram intervalsPediatric cardiologyQTSmartwatchTelemonitoring

Autor:innen und Institutionen

  • Department of Clinical Sciences Lund, Lund University, Lund, Sweden.
  • Department of Pediatric Cardiology, Pediatric Heart Center, Skåne University Hospital, Lund, Sweden.
  • Department of Clinical Sciences Lund, Lund University, Lund, Sweden. Constance.Weismann@med.lu.se.
  • Department of Pediatric Cardiology, Pediatric Heart Center, Skåne University Hospital, Lund, Sweden. Constance.Weismann@med.lu.se.
  • Department of Pediatric Cardiology and Pediatric Intensive Care, Ludwig Maximilium University, Munich, Germany. Constance.Weismann@med.lu.se.

Förderung

  • Lund University

Zitieren

APA

Ernstsson, J., Svensson, B., Liuba, P., & Weismann, C. G. (2024). Validation of smartwatch electrocardiogram intervals in children compared to standard 12 lead electrocardiograms. European Journal of Pediatrics. https://doi.org/10.1007/s00431-024-05648-7

Vancouver

Ernstsson J, Svensson B, Liuba P, Weismann CG. Validation of smartwatch electrocardiogram intervals in children compared to standard 12 lead electrocardiograms. European Journal of Pediatrics. 2024. doi:10.1007/s00431-024-05648-7

BibTeX
@article{Ernstsson2024,
  author = {Ernstsson, Julia and Svensson, Birgitta and Liuba, Petru and Weismann, Constance G},
  title = {Validation of smartwatch electrocardiogram intervals in children compared to standard 12 lead electrocardiograms},
  journal = {European Journal of Pediatrics},
  year = {2024},
  doi = {10.1007/s00431-024-05648-7}
}
Kostenlos starten →