EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieOpen Access · CC BY 4.0

Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control

Lukas O Langanky, Karen B Kreutzer, Christian F Poets, Axel R Franz, Christoph E Schwarz

European Journal of Pediatrics · 2024

DOI: 10.1007/s00431-024-05549-9 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift European Journal of Pediatrics, 2024. Verfasst von Lukas O Langanky und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neonatology, University Children's Hospital, Tübingen University Hospital, Tübingen, Germany. Hauptthemen laut MEDLINE: Oximetry, Cross-Over Studies, Hypoxia, Infant, Premature. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Heidelberg.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO 2 ) measurements during prolonged hypoxemic episodes (pHE) in very preterm infants receiving automatic oxygen control (AOC). We did a post hoc analysis of a randomized crossover study of AOC, programmed to set FiO 2 to “back-up FiO 2 ” during SL. In 24 preterm infants (median (interquartile range)) gestational age 25.3 (24.6 to 25.6) weeks, recording time 12.7 h (12.2 to 13.6 h) per infant, we identified 76 pHEs (median duration 119 s (86 to 180 s)). In 50 (66%) pHEs, SL occurred for a median duration of 51 s (33 to 85 s) and at a median frequency of 2 (1 to 2) SL-periods per pHE. SpO 2 before and after SL was similar (82% (76 to 88%) vs 82% (76 to 87%), p = 0.3)).

Conclusion

SL is common during pHE and must hence be considered in AOC-algorithm designs. Administering a “backup FiO 2 ” (which reflects FiO 2 -requirements during normoxemia) during SL may prolong pHE with SL.

Trial registration

The study was registered at www.clinicaltrials.gov under the registration no. NCT03785899. What is Known: • Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range. What is New: • Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetryCross-Over StudiesHypoxiaInfant, PrematureOxygen SaturationOxygen Inhalation TherapyHumansInfant, NewbornFemaleMaleOxygenInfant, Premature, DiseasesAlgorithms

Stichwörter der Autor:innen

Automated oxygen controlPreterm infantPulse oximetrySignal loss

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Neonatology, University Children's Hospital, Tübingen University Hospital, Tübingen, Germany.
  • Center for Pediatric Clinical Studies, University Children's Hospital, Eberhard Karls University Tübingen, Tübingen, Germany.
  • Department of Neonatology, University Children's Hospital, Tübingen University Hospital, Tübingen, Germany. Christoph.Schwarz@med.uni-heidelberg.de.
  • Department of Neonatology, Center for Pediatric and Adolescent Medicine, University of Heidelberg, Im Neuenheimer Feld 430, 69120, Heidelberg, Germany. Christoph.Schwarz@med.uni-heidelberg.de.

Förderung

  • Universitätsklinikum Heidelberg

Zitieren

APA

Langanky, L. O., Kreutzer, K. B., Poets, C. F., Franz, A. R., & Schwarz, C. E. (2024). Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control. European Journal of Pediatrics. https://doi.org/10.1007/s00431-024-05549-9

Vancouver

Langanky LO, Kreutzer KB, Poets CF, Franz AR, Schwarz CE. Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control. European Journal of Pediatrics. 2024. doi:10.1007/s00431-024-05549-9

BibTeX
@article{Langanky2024,
  author = {Langanky, Lukas O and Kreutzer, Karen B and Poets, Christian F and Franz, Axel R and Schwarz, Christoph E},
  title = {Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control},
  journal = {European Journal of Pediatrics},
  year = {2024},
  doi = {10.1007/s00431-024-05549-9}
}