Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions
European Journal of Applied Physiology · 2024
DOI: 10.1007/s00421-024-05447-w ↗Kurz gefasst
Validierungsstudie in der Zeitschrift European Journal of Applied Physiology, 2024. Verfasst von Harry A Brown und 2 weiteren Autor:innen. Beteiligte Einrichtung: Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia. Hauptthemen laut MEDLINE: Sweat, Exercise, Sodium. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Canberra.
Abstract (Originalsprache)
Purpose
Accurately measuring sweat sodium concentration ([Na + ]) in the field is advantageous for coaches, scientists, and dieticians looking to tailor hydration strategies. The MX3 hydration testing system is a new portable analyser that uses pre-calibrated biosensors to measure sweat [Na + ]. This study aimed to assess the validity and reliability of the MX3 hydration testing system.
Methods
Thirty-one (11 females) recreationally active participants completed one experimental trial. During this trial, participants exercised at a self-selected pace for 45 min in a warm environment (31.5 ± 0.8 °C, 63.2 ± 1.3% relative humidity). Sweat samples were collected from three measurement sites using absorbent patches. The samples were then analysed for sweat [Na + ] using both the MX3 hydration testing system and the Horiba LAQUAtwin-NA-11. The reliability of the MX3 hydration testing system was determined following two measurements of the same sweat sample.
Results
The mean difference between measurements was 0.1 mmoL·L −1 (95% limits of agreement (LoA): − 9.2, 9.4). The analyser demonstrated a coefficient of variation (CV) of 5.6% and the standard error of measurement was 3.3 mmoL·L −1 . When compared to the Horiba LAQUAtwin-NA-11, there was a mean difference of − 1.7 mmoL·L −1 (95% LoA: − 0.25 $${\overline{\text{X}}}$$ X ¯ , 0.25 $${\overline{\text{X}}}$$ X ¯ ) and the CV was 9.8%.
Conclusion
The MX3 hydration testing system demonstrated very good single-trial reliability, moderate agreement and a very good CV relative to the Horiba LAQUAtwin-Na-11. To further validate its performance, the MX3 hydration testing system should be compared with analytical techniques known for superior reliability and validity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia.
- Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia. Julien.Periard@canberra.edu.au.
Förderung
- University of Canberra
Zitieren
APA
Brown, H. A., Clark, B., & Périard, J. D. (2024). Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions. European Journal of Applied Physiology. https://doi.org/10.1007/s00421-024-05447-w
Vancouver
Brown HA, Clark B, Périard JD. Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions. European Journal of Applied Physiology. 2024. doi:10.1007/s00421-024-05447-w
BibTeX
@article{Brown2024,
author = {Brown, Harry A and Clark, Brad and Périard, Julien D},
title = {Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions},
journal = {European Journal of Applied Physiology},
year = {2024},
doi = {10.1007/s00421-024-05447-w}
}
EUVIMED