Long-term safety and efficacy of suprachoroidal ESNOPER CLIP® implantation in non-penetrating deep sclerectomy
Graefe's Archive for Clinical and Experimental Ophthalmology · 14.08.2026
DOI: 10.1007/s00417-026-07396-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Graefe's Archive for Clinical and Experimental Ophthalmology, online erschienen am 14.08.2026. Verfasst von Ignacio Rodriguez-Una und 6 weiteren Autor:innen. Beteiligte Einrichtung: Instituto Oftalmológico Fernández-Vega, Avda. Dres. Fernández-Vega, 34, Oviedo, 33012, Spain. irodriguezu@fernandez-vega.com. Die Arbeit nennt 21 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To evaluate the long-term hypotensive efficacy and safety of the ESNOPER CLIP® implant in non-penetrating deep sclerectomy (NPDS) in patients with open-angle glaucoma. METHODS: Patients who underwent NPDS and suprachoroidal ESNOPER CLIP® implantation were followed up for 5 years. Preoperative and postoperative best corrected visual acuity (BCVA), intraocular pressure (IOP), visual field mean deviation (MD), number of topical hypotensive medications, and intra- and postoperative complications were registered. Surgical success was defined for different IOP thresholds (≤ 21, ≤ 18 and ≤ 15mmHg) with (qualified success) or without postoperative antiglaucoma medications (complete success). RESULTS: 132 eyes (mean age 68.8 ± 8.9 years) were included. BCVA remained similar after surgery (0.73 ± 0.27 decimal preoperatively and 0.74 ± 0.29 after 5 years, p = 0.723). IOP experienced a significant improvement compared to preoperative values (20.7 ± 8.0 vs. 13.0 ± 4.2 mmHg), with a mean number of topical hypotensive medications significantly dropping from 2.63 ± 0.82 to 0.64 ± 0.95 after 5 years. Visual field MD did not significantly change. Four microperforations were registered as intraoperative complications (3%) and 49% needed goniopuncture during follow-up. Needling rate was 26%. After 5 years, complete and qualified success rates were 58% and 88% for IOP≤21mmHg threshold, 59% and 85% for IOP≤18mmHg threshold, and 49% and 60% for IOP≤15mmHg threshold, respectively. CONCLUSIONS: NPDS with the ESNOPER CLIP® implant was associated with sustained reductions in IOP and in the use of topical medications over five years, while maintaining a favourable safety profile. ESNOPER CLIP® implant combined with phacoemulsification appears to be a safe and effective approach for glaucoma patients with a coexisting cataract. KEY MESSAGES: WHAT IS KNOWN : Non-penetrating deep sclerectomy is considered a safe and effective filtering procedure with potentially lower complication rates compared to traditional trabeculectomy. The uveoscleral Esnoper Clip implant used in non-penetrating deep sclerectomy has shown effective and safe results, although available evidence is limited to follow-ups of up to 2 years. WHAT IS NEW: This is the first study reporting 5-year outcomes of the Esnoper Clip implant, including assessment of its performance when combined with cataract surgery. Non-penetrating deep sclerectomy enhanced with the suprachoroidal Esnoper Clip implant is safe and effective in maintaining a reduction of both, intraocular pressure and the number of topical hypotensive medications, in the long term. Non-penetrating deep sclerectomy and Esnoper Clip implantation combined with phacoemulsification is a suitable approach for managing glaucoma patients with concurrent cataract.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Instituto Oftalmológico Fernández-Vega, Avda. Dres. Fernández-Vega, 34, Oviedo, 33012, Spain. irodriguezu@fernandez-vega.com.
- Fundación de Investigación Oftalmológica, University of Oviedo, Oviedo, Spain. irodriguezu@fernandez-vega.com.
- Hospital Universitari Germans Trias i Pujol, Badalona, Spain.
- Centro de Oftalmología Barraquer, Barcelona, Spain.
- Optometry and Vision Department, Faculty of Optics and Optometry, Complutense University of Madrid, Madrid, Spain.
- Clinical and Experimental Eye Research Group, Faculty of Optics and Optometry, Complutense University of Madrid, Madrid, Spain.
Zitieren
APA
Rodriguez-Una, I., Romera-Romero, P., Rodríguez-Carrillo, L., Martínez-Alberquilla, I., Fernandez-Vega-Cueto, A., Rodriguez-Calvo, P. P., & Canut-Jordana, M. I. (2026). Long-term safety and efficacy of suprachoroidal ESNOPER CLIP® implantation in non-penetrating deep sclerectomy. Graefe's Archive for Clinical and Experimental Ophthalmology. https://doi.org/10.1007/s00417-026-07396-w
Vancouver
Rodriguez-Una I, Romera-Romero P, Rodríguez-Carrillo L, Martínez-Alberquilla I, Fernandez-Vega-Cueto A, Rodriguez-Calvo PP, et al. Long-term safety and efficacy of suprachoroidal ESNOPER CLIP® implantation in non-penetrating deep sclerectomy. Graefe's Archive for Clinical and Experimental Ophthalmology. 2026. doi:10.1007/s00417-026-07396-w
BibTeX
@article{RodriguezUna2026,
author = {Rodriguez-Una, Ignacio and Romera-Romero, Pau and Rodríguez-Carrillo, Lola and Martínez-Alberquilla, Irene and Fernandez-Vega-Cueto, Andres and Rodriguez-Calvo, Pedro P and Canut-Jordana, Maria I},
title = {Long-term safety and efficacy of suprachoroidal ESNOPER CLIP® implantation in non-penetrating deep sclerectomy},
journal = {Graefe's Archive for Clinical and Experimental Ophthalmology},
year = {2026},
doi = {10.1007/s00417-026-07396-w}
}
EUVIMED