The influence of electrode array design, scalar dislocation and insertion technique on postoperative vertigo in CI surgery – a prospective study
European Archives of Oto-Rhino-Laryngology · 2025
DOI: 10.1007/s00405-024-09147-1 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Archives of Oto-Rhino-Laryngology, 2025. Verfasst von Manuel Christoph Ketterer und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology, Medical Center, Faculty of Medicine, University of Freiburg, Freiburg, Germany. manuel.christoph.ketterer@uniklini…. Hauptthemen laut MEDLINE: Vertigo, Cochlear Implantation, Postoperative Complications, Cochlear Implants. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Freiburg.
Abstract (Originalsprache)
Objectives
This study aimed to examine the effect of electrode array design, insertion angle, scalar position, and insertion technique on the occurrence of postoperative subjective vertigo following cochlear implant (CI) surgery using questionnaires in conjunction with objective vestibular functional measurements. Materials and methods We prospectively evaluated subjective vertigo using the Dizziness Handicap Inventory (DHI). Additionally, we performed videonystagmography, video head-impulse tests, and vestibular-evoked myogenic potentials to assess the objective vestibular function preoperatively, at four weeks and 12 months after CI. These results were compared with those of postoperative imaging using digital volume tomography.
Results
Postoperative vertigo was observed in 2 out of 62 patients (3%). Cochleostomy ( n = 8) did not lead to an increase in postoperative vertigo. Functional diagnostics revealed abnormalities in up to 23% of patients without subjective dizziness. In our patient cohort, neither electrode array dislocation nor increasing insertion depth was associated with an increase in postoperative vertigo.
Conclusion
Both postoperative vertigo occurrence and electrode array dislocation rates have significantly decreased due to the optimized atraumatic electrode array design and improved surgical insertion techniques. Neither dislocation nor cochleostomy appeared to induce vertigo but the sample size was too small to draw definitive conclusions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otorhinolaryngology, Medical Center, Faculty of Medicine, University of Freiburg, Freiburg, Germany. manuel.christoph.ketterer@uniklinik-freiburg.de.
- Department of Otorhinolaryngology - Head and Neck Surgery, University of Freiburg, Killianstrasse 5, 79106, Freiburg, Germany. manuel.christoph.ketterer@uniklinik-freiburg.de.
- Department of Otorhinolaryngology, Medical Center, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
ORCID-Profile: Manuel Christoph Ketterer
Förderung
- Universitätsklinikum Freiburg
Zitieren
APA
Ketterer, M. C., Rauch, A. K., Beck, R. L., Jakob, T. F., Fries, L., Aschendorff, A., Arndt, S., & Everad, F. (2025). The influence of electrode array design, scalar dislocation and insertion technique on postoperative vertigo in CI surgery – a prospective study. European Archives of Oto-Rhino-Laryngology. https://doi.org/10.1007/s00405-024-09147-1
Vancouver
Ketterer MC, Rauch AK, Beck RL, Jakob TF, Fries L, Aschendorff A, et al. The influence of electrode array design, scalar dislocation and insertion technique on postoperative vertigo in CI surgery – a prospective study. European Archives of Oto-Rhino-Laryngology. 2025. doi:10.1007/s00405-024-09147-1
BibTeX
@article{Ketterer2025,
author = {Ketterer, Manuel Christoph and Rauch, A K and Beck, R L and Jakob, T F and Fries, L and Aschendorff, A and Arndt, S and Everad, F},
title = {The influence of electrode array design, scalar dislocation and insertion technique on postoperative vertigo in CI surgery – a prospective study},
journal = {European Archives of Oto-Rhino-Laryngology},
year = {2025},
doi = {10.1007/s00405-024-09147-1}
}
EUVIMED