Two different methods to digitally visualize continuous electrocochleography potentials during cochlear implantation: a first description of feasibility
European Archives of Oto-Rhino-Laryngology · 2024
DOI: 10.1007/s00405-023-08400-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Archives of Oto-Rhino-Laryngology, 2024. Verfasst von Theda Eichler und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oto-Rhino-Laryngology, Head and Neck Surgery, University Hospital Essen, Essen, Germany. theda.eichler@uk-essen.de. Hauptthemen laut MEDLINE: Cochlear Implantation, Audiometry, Evoked Response, Feasibility Studies, Software. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Essen.
Abstract (Originalsprache)
Purpose
The study explores the potential of real-time electrocochleographic potentials (ECochG) visualization during electrode insertion using digital microscopes such as RoboticScope (BHS®). Collaborative software development of the MAESTRO Software (MED-EL®) offers continuous ECochG monitoring during implantation and postoperative hearing evaluation, addressing previous time constraints. The study aims to assess software applicability and the impact of real-time visualization on long-term residual hearing preservation.
Methods
Eight patients with residual hearing underwent cochlear implantation with Flex26 or Flex28 electrode according to the Otoplan evaluation. ECochG responses were measured and visualized during electrode insertion, with insertion times recorded. Two randomized display methods (graph and arrows) tracked ECochG potentials. Postoperative behavioral thresholds determined hearing preservation. Successful real-time intraoperative ECochG visualization was achieved in all cases, enabling surgeon adaptation. Mean electrode insertion time was 114 s, with postoperative thresholds comparable to preoperative values. Visualization did not affect surgeon workload. ECochG amplitudes differed between patients with and without residual hearing.
Conclusion
The study demonstrates effective implementation of advanced ECochG software combined with real-time visualization, enabling residual hearing preservation during CI. Visualization had no apparent effect on surgeon performance or workload. Future investigation involving a larger population will assess the long-term impact of ECochG on hearing threshold and structure preservation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oto-Rhino-Laryngology, Head and Neck Surgery, University Hospital Essen, Essen, Germany. theda.eichler@uk-essen.de.
- Department of Oto-Rhino-Laryngology, Head and Neck Surgery, University Hospital Essen, Essen, Germany.
Förderung
- Universitätsklinikum Essen
Zitieren
APA
Eichler, T., Lakomek, A., Waschkies, L., Meyer, M., Sadok, N., Lang, S., & Arweiler-Harbeck, D. (2024). Two different methods to digitally visualize continuous electrocochleography potentials during cochlear implantation: a first description of feasibility. European Archives of Oto-Rhino-Laryngology. https://doi.org/10.1007/s00405-023-08400-3
Vancouver
Eichler T, Lakomek A, Waschkies L, Meyer M, Sadok N, Lang S, et al. Two different methods to digitally visualize continuous electrocochleography potentials during cochlear implantation: a first description of feasibility. European Archives of Oto-Rhino-Laryngology. 2024. doi:10.1007/s00405-023-08400-3
BibTeX
@article{Eichler2024,
author = {Eichler, Theda and Lakomek, Antonia and Waschkies, Laura and Meyer, Moritz and Sadok, Nadia and Lang, Stephan and Arweiler-Harbeck, Diana},
title = {Two different methods to digitally visualize continuous electrocochleography potentials during cochlear implantation: a first description of feasibility},
journal = {European Archives of Oto-Rhino-Laryngology},
year = {2024},
doi = {10.1007/s00405-023-08400-3}
}