OPM-based fetal magnetocardiography: fetal cardiac time intervals in healthy pregnancies compared to postnatal ECGs
Archives of Gynecology and Obstetrics · 23.03.2026 · 313 (1)
DOI: 10.1007/s00404-026-08403-5 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Archives of Gynecology and Obstetrics, online erschienen am 23.03.2026. Verfasst von Annette Wacker-Gussmann und 11 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Congenital Heart Disease and Pediatric Cardiology, German Heart Center, TUM University Hospital, Technical University of Munich, Munich, Ge…. Hauptthemen laut MEDLINE: Magnetocardiography, Electrocardiography, Fetal Heart, Heart Rate, Fetal. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Stiftung KinderHerz Deutschland gGmbH, Technische Universität München.
Abstract (Originalsprache)
Background
Fetal magnetocardiography (fMCG) is the most accurate method to assess fetal heart rhythm and conduction. New quantum sensor technology makes it possible to use less expensive devices. The aim of the study is to measure cardiac time intervals of healthy fetuses with a new technology, optically pumped magnetometry (OPM), and compare these results with conventional SQUID-based fMCG and postnatal ECGs.
Methods
The recordings were made using an OPM-based fMCG system and a person-sized magnetic shield, established at German Heart Center,TUM University, Munich, Germany. The subjects were 57 healthy women with uncomplicated singleton pregnancies, studied at a mean gestational age of 32 ± 3.7 weeks with an overall range of 25–40 weeks. The P, PR, QRS, QT, QTc, and RR intervals were measured and compared with published data from previous fMCG devices and postnatal ECG.
Results
The P, PR, and QRS intervals increased with gestational age, but the QT and QTc intervals did not. The measured values of the OPM device were consistent with those from previously published data SQUID values. U-waves were seen in 17.3% of subjects. Eleven subjects were studied by fMCG after 30 weeks’ gestation and by ECG within 17 weeks of birth. In this cohort, the P-wave duration, QRS duration, and QTc increased after birth, but the PR and QT intervals did not.
Conclusion
The results obtained with our innovative OPM-based fMCG system are comparable to previously available measurements obtained by other technologies. The data establish prediction intervals for OPM-based fMCG waveforms in normal fetuses, which is essential for future clinical application. The technology can be used to recognize fetuses with rhythm or conduction abnormalities that might not be evident by echocardiography. To our knowledge, this is the first report comparing fetal cardiac time intervals measured by OPM-based fMCG with postnatal ECG. Lengthening of cardiac intervals consistent with increased chamber size was seen postnatally.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic of Congenital Heart Disease and Pediatric Cardiology, German Heart Center, TUM University Hospital, Technical University of Munich, Munich, Germany. annette.wacker-gussmann@tum.de.
- Clinic of Congenital Heart Disease and Pediatric Cardiology, German Heart Center, TUM University Hospital, Technical University of Munich, Munich, Germany.
- Department of Medical Physics, University of Wisconsin Madison, Madison, WI, USA.
- Departments of Pediatrics (Cardiology) and Biomedical Engineering Medical College of Wisconsin, Milwaukee, WI, USA.
- Chair E66, TUM School of Natural Sciences, Technical University of Munich, Munich, Germany.
- Department of Computer Engineering, Information and Technology, TUM School of Computation, Technical University of Munich, Munich, Germany.
Förderung
- Stiftung KinderHerz Deutschland gGmbH
- Technische Universität München
Zitieren
APA
Wacker-Gussmann, A., Narushima, K., Tardelli, G., Wakai, R. T., Strasburger, J. F., Wunderl, L., Jensch, T., Heckel, R., Lobmaier, S. M., Nagdyman, N., Ewert, P., & Fierlinger, P. (2026). OPM-based fetal magnetocardiography: fetal cardiac time intervals in healthy pregnancies compared to postnatal ECGs. Archives of Gynecology and Obstetrics, 313(1). https://doi.org/10.1007/s00404-026-08403-5
Vancouver
Wacker-Gussmann A, Narushima K, Tardelli G, Wakai RT, Strasburger JF, Wunderl L, et al. OPM-based fetal magnetocardiography: fetal cardiac time intervals in healthy pregnancies compared to postnatal ECGs. Archives of Gynecology and Obstetrics. 2026;313(1). doi:10.1007/s00404-026-08403-5
BibTeX
@article{WackerGussmann2026,
author = {Wacker-Gussmann, Annette and Narushima, Karin and Tardelli, Gabriela and Wakai, Ronald T and Strasburger, Janette F and Wunderl, Lena and Jensch, Tobias and Heckel, Reinhard and Lobmaier, Silvia M and Nagdyman, Nicole and Ewert, Peter and Fierlinger, Peter},
title = {OPM-based fetal magnetocardiography: fetal cardiac time intervals in healthy pregnancies compared to postnatal ECGs},
journal = {Archives of Gynecology and Obstetrics},
year = {2026},
volume = {313},
number = {1},
doi = {10.1007/s00404-026-08403-5}
}
EUVIMED