Ergonomic evaluation of the Senhance® robotic system in minimally invasive gynecologic procedures versus conventional laparoscopy: an exploratory study focusing on surgeon’s muscle activity
Archives of Gynecology and Obstetrics · 16.01.2026 · 313 (1)
DOI: 10.1007/s00404-025-08292-0 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Archives of Gynecology and Obstetrics, online erschienen am 16.01.2026. Verfasst von Bernhard Krämer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Women's Health, University Hospital Tübingen, Calwerstraße 7/6, 72076, Tübingen, Germany. Hauptthemen laut MEDLINE: Ergonomics, Robotic Surgical Procedures, Laparoscopy, Gynecologic Surgical Procedures. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Primary: To evaluate whether robotic-assisted laparoscopic surgery using the Senhance® Surgical System has the potential to reduce muscular demands compared to conventional laparoscopy. Secondary: To verify that the novel eye-tracking feature for zoom and the selection of functions in the Senhance® System menu is not associated with increased eyestrain or neck strain.
Methods
In a within-subject design, 2 experienced surgeons performed 11 robot-assisted and 12 conventional laparoscopic procedures. Muscular demands were monitored throughout surgical procedures by assessing the muscle activity via bipolar surface electromyography of seven muscles of the lower back, shoulder–neck, and forearms. Surgeons’ head, arm, and torso posture was assessed by gravimetrical position sensors. Furthermore, musculoskeletal discomfort, mental and physical workload, task difficulty and working precision were rated by the surgeons. In addition, a ten-item eyestrain questionnaire was administered after each surgical procedure.
Results
Four out of seven muscles were relieved when working with Senhance®. Only in the left shoulder–neck area there was a statistically significant increase in muscle activation associated with robotic-assisted surgery. Changes in surgeons’ posture related to the surgical technique corresponded to the changes in muscle activation. Furthermore, surgeons reported no musculoskeletal discomfort under both conditions and similar levels of workload (mental and physical), and task difficulty. Working precision was subjectively rated to be better during standard laparoscopy. No eyestrain occurred during any of the procedures.
Conclusion
This exploratory study identified the potential of the Senhance® Surgical System for ergonomic improvements and indicated no impairments by the novel eye-tracking feature on muscle demands and eyestrain. Follow-up studies with larger and more diverse indications are needed that also consider clinical outcomes, which were not part of the present study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Women's Health, University Hospital Tübingen, Calwerstraße 7/6, 72076, Tübingen, Germany.
- Institute of Occupational and Social Medicine and Health Services Research, University Hospital Tübingen, University of Tübingen, Wilhelmstraße 27, 72074, Tübingen, Germany.
- Institute of Occupational and Social Medicine and Health Services Research, University Hospital Tübingen, University of Tübingen, Wilhelmstraße 27, 72074, Tübingen, Germany. benjamin.steinhilber@med.uni-tuebingen.de.
Zitieren
APA
Krämer, B., Andress, J., Wohlmeiner, L., Seibt, R., & Steinhilber, B. (2026). Ergonomic evaluation of the Senhance® robotic system in minimally invasive gynecologic procedures versus conventional laparoscopy: an exploratory study focusing on surgeon’s muscle activity. Archives of Gynecology and Obstetrics, 313(1). https://doi.org/10.1007/s00404-025-08292-0
Vancouver
Krämer B, Andress J, Wohlmeiner L, Seibt R, Steinhilber B. Ergonomic evaluation of the Senhance® robotic system in minimally invasive gynecologic procedures versus conventional laparoscopy: an exploratory study focusing on surgeon’s muscle activity. Archives of Gynecology and Obstetrics. 2026;313(1). doi:10.1007/s00404-025-08292-0
BibTeX
@article{Krmer2026,
author = {Krämer, Bernhard and Andress, Jürgen and Wohlmeiner, Leonhard and Seibt, Robert and Steinhilber, Benjamin},
title = {Ergonomic evaluation of the Senhance® robotic system in minimally invasive gynecologic procedures versus conventional laparoscopy: an exploratory study focusing on surgeon’s muscle activity},
journal = {Archives of Gynecology and Obstetrics},
year = {2026},
volume = {313},
number = {1},
doi = {10.1007/s00404-025-08292-0}
}
EUVIMED