Comparison of anatomic axes with a navigated functional rotation axis determined by ligament tension for rotational femoral component alignment in cadaver knee arthroplasty
Archives of Orthopaedic and Trauma Surgery · 2024
DOI: 10.1007/s00402-024-05394-3 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Archives of Orthopaedic and Trauma Surgery, 2024. Verfasst von K Jäckle und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University Medical Center Göttingen, Robert-Koch Str. 40, 37075, Göttingen, Germany. …. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Femur, Cadaver. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Herzzentrum Göttingen.
Abstract (Originalsprache)
Introducion The malimplantation of the total knee arthroplasty (TKA) components is one of the main reasons for revision surgery. For determining the correct intraoperative femoral rotation several anatomic rotational axes were described in order to achieve a parallel, balanced flexion gap. In this cadaveric study prevalent used rotational femoral axes and a navigated functional rotational axis were compared to the flexion-extension axis defined as the gold standard in rotation for femoral TKA component rotation. Materials and methods Thirteen body donors with knee osteoarthritis (mean age: 78.85 ± 6.09; eight females and five males) were examined. Rotational computer tomography was performed on their lower extremities pre- and postoperatively. Knee joint arthroplasties were implanted and CT diagnostics were used to compare the preoperatively determined flexion-extension axis (FEA). The FEA is the axis determined by our surgical technique and serves as an internal reference. It was compared to other axes such as (i) the anatomical transepicondylar axis (aTEA), (ii) the surgical transepicondylar axis (sTEA), (iii) the posterior condylar axis (PCA) and (iv) the functional rotation axis (fRA).
Results
Examination of 26 knee joint arthroplasties revealed a significant angular deviation ( p *** < 0.0001) for all axes when the individual axes and FEA were compared. aTEA show mean angular deviation of 5.2° (± 4.5), sTEA was 2.7° (± 2.2), PCA 2.9° (± 2.3) and the deviation of fRA was 4.3° (± 2.7). A tendency towards external rotation was observed for the relative and maximum axis deviations of the aTEA to the FEA, for the sTEA and the fRA. However, the rotation of the posterior condylar axis was towards inwards.
Conclusions
All axes showed a significant angular deviation from the FEA. We conclude that the presented technique achieves comparable results in terms of FEA reconstruction when compared with the use of the known surrogate axes, with certain deviations in terms of outliers in the internal or external rotation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University Medical Center Göttingen, Robert-Koch Str. 40, 37075, Göttingen, Germany. katharina.jaeckle@med.uni-goettingen.de.
- Department of Trauma Surgery, Orthopaedics and Plastic Surgery, University Medical Center Göttingen, Robert-Koch Str. 40, 37075, Göttingen, Germany.
- Department of Trauma Surgery, University of Magdeburg, Leipziger Str. 44, 39120, Magdeburg, Germany.
- Radiology, Institute for Diagnostic and Interventional Radiology, University Medical Center Göttingen, Robert-Koch Str. 40, 37075, Göttingen, Germany.
Förderung
- Herzzentrum Göttingen
Zitieren
APA
Jäckle, K., Pietzka, M., Schüttrumpf, J. P., Panahi, B., Meier, M. P., Hawellek, T., Lehmann, W., & Walde, T. A. (2024). Comparison of anatomic axes with a navigated functional rotation axis determined by ligament tension for rotational femoral component alignment in cadaver knee arthroplasty. Archives of Orthopaedic and Trauma Surgery. https://doi.org/10.1007/s00402-024-05394-3
Vancouver
Jäckle K, Pietzka M, Schüttrumpf JP, Panahi B, Meier MP, Hawellek T, et al. Comparison of anatomic axes with a navigated functional rotation axis determined by ligament tension for rotational femoral component alignment in cadaver knee arthroplasty. Archives of Orthopaedic and Trauma Surgery. 2024. doi:10.1007/s00402-024-05394-3
BibTeX
@article{Jckle2024,
author = {Jäckle, K and Pietzka, M and Schüttrumpf, J P and Panahi, B and Meier, M-P and Hawellek, T and Lehmann, W and Walde, T A},
title = {Comparison of anatomic axes with a navigated functional rotation axis determined by ligament tension for rotational femoral component alignment in cadaver knee arthroplasty},
journal = {Archives of Orthopaedic and Trauma Surgery},
year = {2024},
doi = {10.1007/s00402-024-05394-3}
}
EUVIMED