Patient specific instrumentation in ACL reconstruction: a proof-of-concept cadaver experiment assessing drilling accuracy when using 3D printed guides
Archives of Orthopaedic and Trauma Surgery · 2023
DOI: 10.1007/s00402-023-05072-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Archives of Orthopaedic and Trauma Surgery, 2023. Verfasst von Mark J M Zee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The N…. Hauptthemen laut MEDLINE: Anterior Cruciate Ligament, Anterior Cruciate Ligament Reconstruction. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Samenwerkingsverband Noord-Nederland.
Abstract (Originalsprache)
Introduction
Accurate positioning of the femoral tunnel in ACL reconstruction is of the utmost importance to reduce the risk of graft failure. Limited visibility during arthroscopy and a wide anatomical variance attribute to femoral tunnel malposition using conventional surgical techniques. The purpose of this study was to determine whether a patient specific 3D printed surgical guide allows for in vitro femoral tunnel positioning within 2 mm of the planned tunnel position. Materials and Methods A patient specific guide for femoral tunnel positioning in ACL reconstruction was created for four human cadaveric knee specimens based on routine clinical MRI data. Fitting properties were judged by two orthopedic surgeons. MRI scanning was performed both pre- and post-procedure. The planned tunnel endpoint was compared to the actual drilled femoral tunnel.
Results
This patient specific 3D printed guide showed a mean deviation of 5.0 mm from the center of the planned femoral ACL origin.
Conclusion
In search to improve accuracy and consistency of femoral tunnel positioning in ACL reconstruction, the use of a patient specific 3D printed surgical guide is a viable option to explore further. The results are comparable to those of conventional techniques; however, further design improvements are necessary to improve accuracy and enhance reproducibility.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedic Surgery, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The Netherlands. m.j.m.zee@umcg.nl.
- 3D Lab, Department of Orthopedic Surgery, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The Netherlands.
- 3D Lab, Department of Oral and Maxillofacial Surgery, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The Netherlands.
- Medical Imaging Center, Department of Radiology, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The Netherlands.
- Department of Orthopedic Surgery, University of Groningen, University Medical Centre Groningen, Hanzeplein 1, PO Box 30.001, 9700RB, Groningen, The Netherlands.
Förderung
- Samenwerkingsverband Noord-Nederland
Zitieren
APA
Zee, M. J. M., Pijpker, P. A. J., Kraeima, J., Viddeleer, A. R., & Diercks, R. L. (2023). Patient specific instrumentation in ACL reconstruction: a proof-of-concept cadaver experiment assessing drilling accuracy when using 3D printed guides. Archives of Orthopaedic and Trauma Surgery. https://doi.org/10.1007/s00402-023-05072-w
Vancouver
Zee MJM, Pijpker PAJ, Kraeima J, Viddeleer AR, Diercks RL. Patient specific instrumentation in ACL reconstruction: a proof-of-concept cadaver experiment assessing drilling accuracy when using 3D printed guides. Archives of Orthopaedic and Trauma Surgery. 2023. doi:10.1007/s00402-023-05072-w
BibTeX
@article{Zee2023,
author = {Zee, Mark J M and Pijpker, Peter A J and Kraeima, Joep and Viddeleer, Alain R and Diercks, Ronald L},
title = {Patient specific instrumentation in ACL reconstruction: a proof-of-concept cadaver experiment assessing drilling accuracy when using 3D printed guides},
journal = {Archives of Orthopaedic and Trauma Surgery},
year = {2023},
doi = {10.1007/s00402-023-05072-w}
}
EUVIMED