Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation
Archives of Orthopaedic and Trauma Surgery · 20.03.2023 · 143 (9) · 5867-5872
DOI: 10.1007/s00402-023-04845-7 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Archives of Orthopaedic and Trauma Surgery, online erschienen am 20.03.2023. Verfasst von Maria Anna Smolle und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Trauma, Medical University of Graz, Graz, Austria. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Osteoarthritis, Knee, Knee Prosthesis. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical University of Graz.
Abstract (Originalsprache)
Introduction
In total knee arthroplasty (TKA), tibial obliquity-restoration using kinematic alignment (KA) poses a major difference to conventional mechanical alignment. This study aimed at analysing the accuracy of conventional instrumentation (CI) versus patient-specific instrumentation (PSI) to restore anatomic tibial obliquity measured by the medial proximal tibial angle (MPTA) on conventional X-rays. Materials and Methods One-hundred patients were randomized to receive CI ( n = 50) or PSI ( n = 50) for TKA. Further 100 patients received CI without randomisation, resulting in 200 patients in total (127 women, mean age: 70.7 (range: 48–90 years). Pre- and postoperative X-rays were measured twice by two observers with a 2-week break in-between. Inter- and intraclass correlations were calculated and postoperative tibial obliquity compared to preoperative anatomy.
Results
In 150 patients with CI, no case with tibial obliquity-deviation greater than 2° was found, whilst 21.3% ( n = 32) and 0.7% ( n = 1) of cases and had a deviation of 0°–1°, and 1°–2°, respectively. In the remaining 78.0% ( n = 117), tibial obliquity was restored. In 50 patients with PSI, no single case with a deviation greater than 1° was found. Sixty percent ( n = 30) had a deviation of 0°–1°. In the remaining 40.0% ( n = 20), no deviation from preoperative measurements was found. Consequently, CI resulted in a significantly smaller change in tibial obliquity from preoperative to postoperative than PSI ( p < 0.001). Inter- and intra-class correlations showed a substantial agreement (any ICC > 0.90).
Conclusion
Both conventional and patient-specific instrumentation revealed adequate results with respect to restoring tibial obliquity in kinematically aligned TKA, with conventional instrumentation achieving superior results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics and Trauma, Medical University of Graz, Graz, Austria.
- Department of Orthopaedics and Trauma, Medical University of Graz, Graz, Austria. Patrick.sadoghi@medunigraz.at.
Förderung
- Medical University of Graz
Zitieren
APA
Smolle, M. A., Koutp, A., Clar, C., Leitner, L., Leithner, A., & Sadoghi, P. (2023). Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation. Archives of Orthopaedic and Trauma Surgery, 143(9), 5867-5872. https://doi.org/10.1007/s00402-023-04845-7
Vancouver
Smolle MA, Koutp A, Clar C, Leitner L, Leithner A, Sadoghi P. Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation. Archives of Orthopaedic and Trauma Surgery. 2023;143(9):5867-5872. doi:10.1007/s00402-023-04845-7
BibTeX
@article{Smolle2023,
author = {Smolle, Maria Anna and Koutp, Amir and Clar, Clemens and Leitner, Lukas and Leithner, Andreas and Sadoghi, Patrick},
title = {Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation},
journal = {Archives of Orthopaedic and Trauma Surgery},
year = {2023},
volume = {143},
number = {9},
pages = {5867--5872},
doi = {10.1007/s00402-023-04845-7}
}
EUVIMED