Impact of powered circular stapling devices on anastomotic leakage rates in colorectal surgery
International Journal of Colorectal Disease · 08.07.2026 · 41 (1)
DOI: 10.1007/s00384-026-05195-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Colorectal Disease, online erschienen am 08.07.2026. Verfasst von Catherine Kollmann und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of General, Visceral, Transplant, Vascular and Paediatric Surgery, University Hospital Würzburg, Oberdürrbacherstrasse 6, 97080, Würzburg,…. Hauptthemen laut MEDLINE: Anastomotic Leak, Surgical Staplers, Surgical Stapling. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Würzburg.
Abstract (Originalsprache)
Introduction
Anastomotic leakage (AL) remains a severe complication after colorectal surgery, increasing morbidity and mortality. Powered circular staplers may influence anastomotic healing through standardised compression and power distribution. This study aimed to compare clinical outcomes of powered versus manual circular staplers for transanal anastomoses.
Method
This retrospective single-centre cohort study included consecutive patients undergoing colorectal resection with transanal circular anastomosis between January 2022 and June 2025. Primary endpoint was the incidence of AL according to stapler type. To ensure comparability between groups, propensity score matching was performed.
Results
Among 333 patients, powered staplers were used in 69.1% and manual staplers in 30.9%. The overall AL rate was 15.0%. AL was significantly associated with higher BMI (28.1 vs. 26.6 kg/m²; p = .021), ASA ≥ 3 (64.0 vs. 46.0%; p = .019), diabetes mellitus (24.0 vs. 8.1%; p .001), open (34.0 vs. 19.1%) or converted surgeries (14.0 vs. 4.9%; p = .004) and greater intraoperative fluid administration (2548.1 vs. 2188.8 ml; p = .015). Most patients (72.1%) followed enhanced recovery concepts (FAST TRACK) with trend toward lower adherence in the AL cohort (n.s.). After propensity score matching, AL occurred in 9.7% following powered circular stapling and 11.8% following manual circular stapling (n.s.).
Conclusion
In this cohort, powered stapling devices were not associated with lower anastomotic leakage rates compared with manual staplers for transanal circular anastomosis in patients managed under modern minimally invasive and FAST TRACK concepts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of General, Visceral, Transplant, Vascular and Paediatric Surgery, University Hospital Würzburg, Oberdürrbacherstrasse 6, 97080, Würzburg, Germany.
- Department of General, Visceral, Transplant, Vascular and Paediatric Surgery, University Hospital Würzburg, Oberdürrbacherstrasse 6, 97080, Würzburg, Germany. Flemming_S@ukw.de.
ORCID-Profile: Sven Flemming
Förderung
- Universitätsklinikum Würzburg
Zitieren
APA
Kollmann, C., Eckart, T., Kusnezov, B., Kollmann, L., Kelm, M., Germer, C. T., Lock, J. F., & Flemming, S. (2026). Impact of powered circular stapling devices on anastomotic leakage rates in colorectal surgery. International Journal of Colorectal Disease, 41(1). https://doi.org/10.1007/s00384-026-05195-7
Vancouver
Kollmann C, Eckart T, Kusnezov B, Kollmann L, Kelm M, Germer CT, et al. Impact of powered circular stapling devices on anastomotic leakage rates in colorectal surgery. International Journal of Colorectal Disease. 2026;41(1). doi:10.1007/s00384-026-05195-7
BibTeX
@article{Kollmann2026,
author = {Kollmann, Catherine and Eckart, Theresa and Kusnezov, Beata and Kollmann, Lars and Kelm, Matthias and Germer, Christoph-Thomas and Lock, Johan Friso and Flemming, Sven},
title = {Impact of powered circular stapling devices on anastomotic leakage rates in colorectal surgery},
journal = {International Journal of Colorectal Disease},
year = {2026},
volume = {41},
number = {1},
doi = {10.1007/s00384-026-05195-7}
}
EUVIMED