Drill choice as a modifiable factor in pediatric sEEG safety
Child's Nervous System · 20.06.2026 · 42 (1)
DOI: 10.1007/s00381-026-07319-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift Child's Nervous System, online erschienen am 20.06.2026. Verfasst von Mahalia Dalmage und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Biological Sciences, University of Chicago Pritzker School of Medicine, Chicago, IL, USA. Hauptthemen laut MEDLINE: Stereotaxic Techniques, Electroencephalography, Neurosurgical Procedures. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Stereoelectroencephalography (sEEG) procedures require drilling serial burr holes for electrode placement. Key drilling characteristics include haptic feedback, ease of use, and time-efficient burr hole creation. Orthopedic drill bits are often repurposed for sEEG. We present a case series highlighting sEEG as an application for a new-to-market neurosurgery-tailored drill.
Methods
We performed a retrospective chart review on patients who underwent sEEG at our institution from 2022–2023. Patients were grouped by type of drill, Phasor ® (a disposable battery-operated drill 3.20-mm diameter, 210-mm usable length) or Stryker ® (Orthopedic Drill Cordless).
Results
Twelve patients underwent sEEG with the neurosurgery-tailored drill (2 unilateral and 10 bilateral); on average 10.7 ± 1.6 electrodes were placed per patient and mean operative duration was 162.9 ± 49 min. Tactile feel of bony layers and consistent drill time per burr hole (< 15 s) was subjectively noted. No complications were noted. Five patients underwent sEEG with the Stryker ® orthopedic drill (2 unilateral and 3 bilateral); an average of 11 ± 2.9 electrodes were placed per patient, mean operative duration was 176.4 ± 18.9 min. Subjectively, drill time per burr hole was variable and increased in duration between the first and final hole. One patient incurred a skin burn at the drill site which led to wound dehiscence and infection requiring prolonged wound care and antibiotics.
Conclusion
The Phasor ® drill offers a lightweight, reliable, consistent, safe and efficient option for sEEG burr hole creation and may be considered as an alternative to the Stryker ® orthopedic drill.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Biological Sciences, University of Chicago Pritzker School of Medicine, Chicago, IL, USA.
- Division of Pediatric Neurosurgery, Ann and Robert H. Lurie Children's Hospital, Chicago, IL, USA.
- Department of Neurosurgery, Northwestern University Feinberg School of Medicine, Chicago, IL, USA.
- Department of Neurological Surgery, Rush University Medical Center, Chicago, IL, USA.
- Marshall University Joan C. Edwards School of Medicine, Huntington, WV, USA.
- Division of Pediatric Neurosurgery, Ann and Robert H. Lurie Children's Hospital, Chicago, IL, USA. jraskin@luriechildrens.org.
ORCID-Profile: Oludamilola Adeshina, Jeffrey S Raskin
Zitieren
APA
Dalmage, M., Abdelmageed, S., Chiu, L., Adeshina, O., Poland, E., & Raskin, J. S. (2026). Drill choice as a modifiable factor in pediatric sEEG safety. Child's Nervous System, 42(1). https://doi.org/10.1007/s00381-026-07319-4
Vancouver
Dalmage M, Abdelmageed S, Chiu L, Adeshina O, Poland E, Raskin JS. Drill choice as a modifiable factor in pediatric sEEG safety. Child's Nervous System. 2026;42(1). doi:10.1007/s00381-026-07319-4
BibTeX
@article{Dalmage2026,
author = {Dalmage, Mahalia and Abdelmageed, Sunny and Chiu, Lucinda and Adeshina, Oludamilola and Poland, Emma and Raskin, Jeffrey S},
title = {Drill choice as a modifiable factor in pediatric sEEG safety},
journal = {Child's Nervous System},
year = {2026},
volume = {42},
number = {1},
doi = {10.1007/s00381-026-07319-4}
}