The dynamic response of a pressure transducer for measurements in water

M. Jentzsch, W. Lechner, R. Woszidlo, C. N. Nayeri, C. O. Paschereit

Experiments in Fluids · 2024-08 · 65 (8)

DOI: 10.1007/s00348-024-03854-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Experiments in Fluids, online erschienen am 14.08.2024. Verfasst von M. Jentzsch und 4 weiteren Autor:innen. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Technische Universität Berlin.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Dynamic pressure measurements are indispensable in the field of fluid mechanics. Attaching tubing as a transmission line to the pressure transducer is often unavoidable but significantly reduces the usable bandwidth of the measurement system. Complex fluid-wall interactions and potential outgassing of air are present within systems with water-filled tubes. Comprehensive studies aiding researchers in selecting suitable transmission line parameters (i.e., material, length, and diameter) are not available. A simple calibration apparatus is designed for the frequency response characterization of multiple pressure transducers simultaneously applying a pressure step. The setup is thoroughly characterized and a detailed description is provided to optimize the bandwidth. A piezoresistive pressure transducer attached to water-filled tubes, as commonly used in hydrodynamic experiments, is characterized in the low-frequency range (i.e., $$f \le {300}$$ f ≤ 300 Hz). Tube-related effects, such as length, diameter, and material are investigated. The impact of entrapped air within the tubing is analyzed. The feasibility of substituting water with silicone oil to fill the tubes is explored. To optimize the usable bandwidth of the pressure measurement system for dynamic applications, it is essential to maintain short tubing that is as rigid as possible and free from entrapped air. Pressure wave propagation speed is reduced by two orders of magnitude in elastic transmission lines made of silicone. Pressure corrections through dynamic calibration are challenging due to the system’s sensitivity to various parameters affecting the dynamic response.

Förderung

  • Technische Universität Berlin

Zitieren

APA

Jentzsch, M., Lechner, W., Woszidlo, R., Nayeri, C. N., & Paschereit, C. O. (2024). The dynamic response of a pressure transducer for measurements in water. Experiments in Fluids, 65(8). https://doi.org/10.1007/s00348-024-03854-5

Vancouver

Jentzsch M, Lechner W, Woszidlo R, Nayeri CN, Paschereit CO. The dynamic response of a pressure transducer for measurements in water. Experiments in Fluids. 2024;65(8). doi:10.1007/s00348-024-03854-5

BibTeX
@article{Jentzsch2024,
  author = {Jentzsch, M. and Lechner, W. and Woszidlo, R. and Nayeri, C. N. and Paschereit, C. O.},
  title = {The dynamic response of a pressure transducer for measurements in water},
  journal = {Experiments in Fluids},
  year = {2024},
  volume = {65},
  number = {8},
  doi = {10.1007/s00348-024-03854-5}
}
Kostenlos starten →