Optical characterization of endoscopes for StoneSense™ compatibility

Nicola Nannola iD, Carlos Gonzalez-Gonzalez, Federico Zorzi, Pietro Scilipoti, Adèle Martin-Illac, Frédéric Panthier, Marie Chicaud, Steeve Doizi, Iliya Yaroslavsky, Anastasia Kovalenko, James Childs, Igor Perchuk und 1 weitere

World Journal of Urology · 02.11.2026 · 44 (1)

DOI: 10.1007/s00345-026-06845-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift World Journal of Urology, online erschienen am 02.11.2026. Verfasst von Nicola Nannola und 12 weiteren Autor:innen. Beteiligte Einrichtung: Service d'urologie, Tenon Hôpital, 4 Rue de la Chine, 75020, Paris, France. nicolanannola01@gmail.com. Hauptthemen laut MEDLINE: Ureteroscopes, Lithotripsy, Laser, Cystoscopes, Ureteroscopy. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and objective The StoneSense Module (SSM), integrated within IPG laser lithotripsy systems, uses endoscopic white-light illumination for real-time stone-tissue discrimination, enabling conditional laser firing only in the presence of a confirmed calculus. SSM performance depends on the endoscope’s optical output falling within defined thresholds. We aimed to characterize commercially available flexible ureteroscopes and cystoscopes against predefined SSM compatibility criteria.

Methods

Twenty device-videobox combinations (15 flexible ureteroscopes, 4 cystoscopes, 1 semirigid ureteroscope; 7 manufacturers) with complete StoneSense-processed band data were included. Total power and spectral band power (Bands 1–3) were measured at three brightness levels in free-space configuration (no fiber inserted), and temporal profile was classified by oscilloscopy. Per-fiber band power was modeled from these free-space measurements, not measured with the fiber in the working channel. Predicted SSM band-power compatibility was assessed per device-brightness-fiber combination against predefined thresholds, with a minimum PWM frequency of 2 kHz as the temporal criterion.

Results

All 20 devices exhibited continuous-wave (CW) emission, conferring unconditional temporal SSM compatibility. Following recalibration of the SSM reference thresholds, the 200 µm and 365 µm fibers were predicted to meet SSM band power thresholds at medium/maximum brightness for 18 of the 20 devices, with the Karl Storz Flex XC and the Olympus CYF-VH cystoscope as the two exceptions at both fiber sizes. Additional saturation, restricted to the 550 µm and 940 µm fibers, was identified for six further devices (Table 1). Four Olympus platforms showed modeled sub-threshold power at minimum brightness. Four illumination archetypes were identified: blue-pump LED, red-shifted LED, Olympus intermediate LED, and LED-phosphor.

Conclusions

The 200 µm and 365 µm fibers provide the broadest predicted compatibility (18/20 platforms), with the Flex XC and CYF-VH as the exceptions. Predicted incompatibilities are otherwise device-architecture-driven and manageable through fiber and brightness selection. A device-level compatibility reference matrix based on modeled estimates is provided.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

UreteroscopesLithotripsy, LaserCystoscopesUreteroscopyEquipment DesignHumans

Stichwörter der Autor:innen

CystoscopyFlexible ureteroscopyLaser lithotripsyOptical compatibilityPulse-width modulationSpectrophotometryStone recognitionStoneSense

Autor:innen und Institutionen

  • Service d'urologie, Tenon Hôpital, 4 Rue de la Chine, 75020, Paris, France. nicolanannola01@gmail.com.
  • GRC20 - Endolase lab, Sorbonne University, PIMM-Arts et Métiers Paris Tech, F-75013, Paris, France. nicolanannola01@gmail.com.
  • PIMM, UMR 8006 CNRS-Arts et Métiers ParisTech, 151 bd de l'Hôpital, F-75013, Paris, France. nicolanannola01@gmail.com.
  • Department of Urology, University of Modena and Reggio Emilia, 41121, Modena, Italy. nicolanannola01@gmail.com.
  • Service d'urologie, Tenon Hôpital, 4 Rue de la Chine, 75020, Paris, France.
  • GRC20 - Endolase lab, Sorbonne University, PIMM-Arts et Métiers Paris Tech, F-75013, Paris, France.

ORCID-Profile: Nicola Nannola

Zitieren

APA

Nannola, N., Gonzalez-Gonzalez, C., Zorzi, F., Scilipoti, P., Martin-Illac, A., Panthier, F., Chicaud, M., Doizi, S., Yaroslavsky, I., Kovalenko, A., Childs, J., Perchuk, I., & Traxer, O. (2026). Optical characterization of endoscopes for StoneSense™ compatibility. World Journal of Urology, 44(1). https://doi.org/10.1007/s00345-026-06845-9

Vancouver

Nannola N, Gonzalez-Gonzalez C, Zorzi F, Scilipoti P, Martin-Illac A, Panthier F, et al. Optical characterization of endoscopes for StoneSense™ compatibility. World Journal of Urology. 2026;44(1). doi:10.1007/s00345-026-06845-9

BibTeX
@article{Nannola2026,
  author = {Nannola, Nicola and Gonzalez-Gonzalez, Carlos and Zorzi, Federico and Scilipoti, Pietro and Martin-Illac, Adèle and Panthier, Frédéric and Chicaud, Marie and Doizi, Steeve and Yaroslavsky, Iliya and Kovalenko, Anastasia and Childs, James and Perchuk, Igor and Traxer, Olivier},
  title = {Optical characterization of endoscopes for StoneSense™ compatibility},
  journal = {World Journal of Urology},
  year = {2026},
  volume = {44},
  number = {1},
  doi = {10.1007/s00345-026-06845-9}
}
Kostenlos starten →