Übersichtsarbeit in der Zeitschrift World Journal of Urology, online erschienen am 24.05.2023. Verfasst von Joao G Porto und 7 weiteren Autor:innen. Beteiligte Einrichtung: Desai Sethi Urology Institute, University of Miami, Miller School of Medicine, Coral Gables, USA. Hauptthemen laut MEDLINE: Lower Urinary Tract Symptoms, Prostatic Hyperplasia, Transurethral Resection of Prostate. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 30 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: MAUDE database houses medical device reports of suspected device-related complications received by Food and Drug Administration. In the present study we aim to evaluate the MAUDE database for reported complications of MIST procedures. METHODS: The database was queried using keywords: rezum, urolift, prostate embolization (PAE), transurethral needle ablation (TUNA), transurethral microwave therapy (TUMT), prostate stent and Temporarily Implanted Nitinol Device (iTIND) on 10/1/22 to extract information regarding device problems and procedure-related complications. Gupta classification system was used to stratify complications. Statistical analysis was performed to compare frequency of complications among MIST procedures. RESULTS: We found a total of 692 reports (Rezum-358, urolift-226, PAE-53, TUNA-31, TUMT-19, prostatic stent-4, and iTIND-1). Most complications related to device or users were minor (level 1 and 2) and there was no significant difference among various MIST procedures. The screen/system error was responsible for 93% and 83% aborted cases in Rezum and TUNA, respectively, and PAE showed 40% of device component detachment/fracture. Overall Urolift and TUMT were associated with statistically significant higher incidence of major (level 3 and 4) complications (23% and 21%, respectively) as compared with Rezum (7%). Most major complications needing hospitalization after Urolift included hematoma and hematuria with clots and those after Rezum included urinary tract infection and sepsis. Thirteen deaths were reported, mostly due to cardiovascular events, which were classified as not associated with the proposed treatment. CONCLUSION: MIST for BPH can occasionally cause significant morbidity. Our data should assist urologists and patients in shared decision-making process.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lower Urinary Tract SymptomsProstatic HyperplasiaTransurethral Resection of ProstateHumansMaleMinimally Invasive Surgical ProceduresProstateProstatectomyTreatment Outcome

Stichwörter der Autor:innen

Benign prostatic hyperplasiaComplicationsEnlarged prostateMinimally invasive surgical therapyRezumUrolift

Autor:innen und Institutionen

  • Desai Sethi Urology Institute, University of Miami, Miller School of Medicine, Coral Gables, USA.
  • Seth GS Medical College and KEM Hospital, Mumbai, India.
  • Department of Interventional Radiology, University of Miami, Miller School of Medicine, Coral Gables, USA.
  • Desai Sethi Urology Institute, University of Miami, Miller School of Medicine, Coral Gables, USA. hns35@miami.edu.

Zitieren

APA

Porto, J. G., Arbelaez, M. C. S., Blachman-Braun, R., Bhatia, A., Bhatia, S., Satyanarayana, R., Marcovich, R., & Shah, H. N. (2023). Complications associated with minimally invasive surgical therapies (MIST) for surgical management of benign prostatic hyperplasia: a Manufacturer and User Facility Device Experience (MAUDE) database review. World Journal of Urology, 41(7), 1975-1982. https://doi.org/10.1007/s00345-023-04440-w

Vancouver

Porto JG, Arbelaez MCS, Blachman-Braun R, Bhatia A, Bhatia S, Satyanarayana R, et al. Complications associated with minimally invasive surgical therapies (MIST) for surgical management of benign prostatic hyperplasia: a Manufacturer and User Facility Device Experience (MAUDE) database review. World Journal of Urology. 2023;41(7):1975-1982. doi:10.1007/s00345-023-04440-w

BibTeX
@article{Porto2023,
  author = {Porto, Joao G and Arbelaez, Maria Camila Suarez and Blachman-Braun, Ruben and Bhatia, Ansh and Bhatia, Shivank and Satyanarayana, Ramgopal and Marcovich, Robert and Shah, Hemendra N},
  title = {Complications associated with minimally invasive surgical therapies (MIST) for surgical management of benign prostatic hyperplasia: a Manufacturer and User Facility Device Experience (MAUDE) database review},
  journal = {World Journal of Urology},
  year = {2023},
  volume = {41},
  number = {7},
  pages = {1975--1982},
  doi = {10.1007/s00345-023-04440-w}
}
Kostenlos starten →