Diagnostic assessment of artificial intelligence reconstruction on accelerated prostate MRI: a retrospective, paired, multi-reader multi-case study
European Radiology · 02.04.2026 · 36 (8) · 6440-6449
DOI: 10.1007/s00330-026-12479-7 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift European Radiology, online erschienen am 02.04.2026. Verfasst von Quintin van Lohuizen und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands. q.y.van.lohuizen@umcg.nl. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Magnetic Resonance Imaging, Artificial Intelligence, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, Health~Holland and Siemens Healthineers.
Abstract (Originalsprache)
Objectives
To determine whether AI-reconstructed prostate MRI at reduced acquisition times maintains prostate cancer (PCa) detection performance comparable to conventional scans. Materials and methods This multicenter, retrospective, consecutive-cohort study included 120 multi-coil T2-weighted prostate MRI scans from the University Medical Center Groningen (UMCG) and 312 publicly available scans from New York University (NYU). An AI model trained on the NYU data was tested on retrospectively undersampled UMCG scans at acceleration factors R = 3 and R = 6 (i.e., data reduction in k-space). Eight experienced radiologists participated in a multi-reader multi-case PCa detection study. Diagnostic performance was assessed using the area under the receiver operating characteristic curve (AUROC). Histopathology and PI-RADS ≤ 2 findings served as reference standards. Multiple image quality metrics were subjectively evaluated using a 4-point Likert scale.
Results
No statistically significant reduction in PCa detection was observed at an MRI acceleration up to R = 6 ( p = 0.08). AUROC values were 0.86 (95% CI: 0.74–0.90) for R = 1, 0.82 (0.72–0.88) for R = 3, and 0.80 (0.70–0.86) for R = 6. Compared to R = 1, R = 3 scans were rated by radiologists to have significantly improved sharpness (+0.2, p < 0.05) and lower noise (+0.1, p < 0.05). Overall visual quality at R = 6 remained comparable to R = 1 (2.81 at R6 vs. 2.74 at R1).
Conclusion
AI-driven reconstruction enabled a sixfold acceleration of T2-weighted prostate MRI (0:33–1:27 min) without a statistically significant reduction in PCa detection, while preserving perceived image quality. However, the decreasing diagnostic performance at higher accelerations warrants further prospective evaluation. Key Points Question This study investigated whether deep learning reconstruction enables three- to sixfold acceleration without reducing radiologists’ detection of clinically significant prostate cancer .
Findings
In a multi-reader multi-case study with eight radiologists, three- and sixfold acceleration showed no significant change in area under the receiver operating characteristic curve . Clinical relevance Deep learning reconstruction shortened T2-weighted acquisition times at sixfold acceleration while preserving perceived image quality and diagnostic performance across acceleration factors .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands. q.y.van.lohuizen@umcg.nl.
- Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands. s.j.fransen@umcg.nl.
- Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, The Netherlands.
- Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
- Department of Radiology, University Medical Center Groningen, Groningen, The Netherlands.
- Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Förderung
- NCI NIH HHS
- Health~Holland and Siemens Healthineers
Zitieren
APA
van Lohuizen, Q., Fransen, S. J., Yiasemis, G., Twilt, J. J., Roest, C., Arita, Y., Borstlap, J., Fütterer, J. J., de Rooij, M., Rouw, D. B., Schoots, I. G., Turkbey, B., Withey, S. J., Simonis, F. F. J., Huisman, H., Kwee, T. C., Teuwen, J., & Yakar, D. (2026). Diagnostic assessment of artificial intelligence reconstruction on accelerated prostate MRI: a retrospective, paired, multi-reader multi-case study. European Radiology, 36(8), 6440-6449. https://doi.org/10.1007/s00330-026-12479-7
Vancouver
van Lohuizen Q, Fransen SJ, Yiasemis G, Twilt JJ, Roest C, Arita Y, et al. Diagnostic assessment of artificial intelligence reconstruction on accelerated prostate MRI: a retrospective, paired, multi-reader multi-case study. European Radiology. 2026;36(8):6440-6449. doi:10.1007/s00330-026-12479-7
BibTeX
@article{vanLohuizen2026,
author = {van Lohuizen, Quintin and Fransen, Stefan Johannes and Yiasemis, George and Twilt, Jasper Jonathan and Roest, Christian and Arita, Yuki and Borstlap, Jacob and Fütterer, Jurgen J and de Rooij, Maarten and Rouw, Dennis B and Schoots, Ivo G and Turkbey, Baris and Withey, Samuel Joseph and Simonis, Frank F J and Huisman, Henkjan and Kwee, Thomas Christian and Teuwen, Jonas and Yakar, Derya},
title = {Diagnostic assessment of artificial intelligence reconstruction on accelerated prostate MRI: a retrospective, paired, multi-reader multi-case study},
journal = {European Radiology},
year = {2026},
volume = {36},
number = {8},
pages = {6440--6449},
doi = {10.1007/s00330-026-12479-7}
}