Optimization of measurements with an ultrasound attenuation coefficient algorithm for quantifying liver fat
European Radiology · 2025
DOI: 10.1007/s00330-025-11906-5 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift European Radiology, 2025. Verfasst von Giovanna Ferraioli und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, Campus della Salute, Pavia, Italy. Hauptthemen laut MEDLINE: Algorithms, Liver, Adipose Tissue, Fatty Liver. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Methods for measuring the ultrasound attenuation coefficient (AC) vary across different systems. Some have fixed regions of interest (ROI) while others have movable ROIs. Aims were to evaluate whether, using a system with a fixed ROI, correlation between AC and MRI proton density fat fraction (MRI-PDFF), and performance could be improved by (i) reducing fixed ROI length to 30 mm, changing starting point from the transducer, and (ii) using a movable ROI at different depths. Materials and methods In this retrospective multicenter study, AC measurements were performed with the Arietta 850 system, and raw data were automatically stored. AC values were obtained using a standard commercial algorithm (ROI-setting1, 35–75 mm from transducer). Raw data were successively reprocessed externally using a fixed 45–75 mm ROI (ROI-setting2) and a movable 30 mm ROI positioned with the top at 20 mm (ROI-setting3) and 25 mm below the liver capsule (ROI-setting4). Spearman rho and area under the receiver operating characteristics curve (AUROC) were used to assess correlation with MRI-PDFF and performance, respectively, and the Delong test was used to compare AUROCs.
Results
Seven hundred fifty participants (median age: 65 [52, 73] years; 384 males) were included. Correlation of ROI-setting1 with MRI-PDFF was 0.75 (0.72, 0.78), reaching 0.80 (0.77, 0.82) with ROI-setting4. Overall, ROI-setting4 showed significantly the best performance across steatosis grades. AUROCs for S > 0 were 0.90 (0.87, 0.92) for ROI-setting1 and 0.92 (0.90, 0.94) with ROI-setting4 ( p < 0.001). This latter performed significantly better than all other settings in participants with obesity and skin-to-liver distance > 25 mm ( p < 0.05).
Conclusions
A movable ROI improves both AC correlation with MRI-PDFF and performance. The highest improvement was with the ROI top 25 mm below the liver capsule. Key Points Question It remains uncertain whether utilizing an ROI at a fixed position from the transducer is optimal for ultrasound AC measurements .
Findings
The highest AC performance and correlation with MRI-PDFF were observed with a movable ROI positioned with the top at 25 mm under the liver capsule . Clinical relevance A movable ROI with consistent sub-capsular placement improves both correlation between AC and MRI-PDFF, as well as its performance, whereas in individuals with a skin-to-liver capsule distance > 25 mm, a fixed ROI position significantly decreases AC performance . Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, Campus della Salute, Pavia, Italy.
- SOD Medicina Interna ed Epatologia, Azienda Ospedaliero-Universitaria Careggi, Firenze, Italy.
- Sheila Sherlock Liver Unit and UCL Institute for Liver and Digestive Health, Royal Free Hospital, London, UK.
- Department of Nursing, Faculty of Nursing, Gifu Kyoritsu University, Ogaki, Japan.
- Department of Imaging Diagnosis, Ogaki Municipal Hospital, Ogaki, Japan.
- Department of Gastroenterology and Hepatology, Suita Municipal Hospital, Suita, Japan.
Zitieren
APA
Ferraioli, G., Roccarina, D., Kumada, T., Ogawa, S., Yoshida, Y., Hirooka, M., & Barr, R. G. (2025). Optimization of measurements with an ultrasound attenuation coefficient algorithm for quantifying liver fat. European Radiology. https://doi.org/10.1007/s00330-025-11906-5
Vancouver
Ferraioli G, Roccarina D, Kumada T, Ogawa S, Yoshida Y, Hirooka M, et al. Optimization of measurements with an ultrasound attenuation coefficient algorithm for quantifying liver fat. European Radiology. 2025. doi:10.1007/s00330-025-11906-5
BibTeX
@article{Ferraioli2025,
author = {Ferraioli, Giovanna and Roccarina, Davide and Kumada, Takashi and Ogawa, Sadanobu and Yoshida, Yuichi and Hirooka, Masashi and Barr, Richard G},
title = {Optimization of measurements with an ultrasound attenuation coefficient algorithm for quantifying liver fat},
journal = {European Radiology},
year = {2025},
doi = {10.1007/s00330-025-11906-5}
}
EUVIMED