Concordance between single-slice abdominal computed tomography-based and bioelectrical impedance-based analysis of body composition in a prospective study
European Radiology · 19.06.2025 · 35 (12) · 8000-8011
DOI: 10.1007/s00330-025-11746-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Radiology, online erschienen am 19.06.2025. Verfasst von Uli Fehrenbach und 9 weiteren Autor:innen. Beteiligte Einrichtung: Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Radiology, Berlin,…. Hauptthemen laut MEDLINE: Electric Impedance, Body Composition, Tomography, X-Ray Computed. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Body composition analysis (BCA) is a recognized indicator of patient frailty. Apart from the established bioelectrical impedance analysis (BIA), computed tomography (CT)-derived BCA is being increasingly explored. The aim of this prospective study was to directly compare BCA obtained from BIA and CT. Materials and methods A total of 210 consecutive patients scheduled for CT, including a high proportion of cancer patients, were prospectively enrolled. Immediately prior to the CT scan, all patients underwent BIA. CT-based BCA was performed using a single-slice AI tool for automated detection and segmentation at the level of the third lumbar vertebra (L3). BIA-based parameters, body fat mass (BFM BIA ) and skeletal muscle mass (SMM BIA ), CT-based parameters, subcutaneous and visceral adipose tissue area (SATA CT and VATA CT ) and total abdominal muscle area (TAMA CT ) were determined. Indices were calculated by normalizing the BIA and CT parameters to patient’s weight (body fat percentage (BFP BIA ) and body fat index (BFI CT )) or height (skeletal muscle index (SMI BIA ) and lumbar skeletal muscle index (LSMI CT )).
Results
Parameters representing fat, BFM BIA and SATA CT + VATA CT , and parameters representing muscle tissue, SMM BIA and TAMA CT , showed strong correlations in female (fat: r = 0.95; muscle: r = 0.72; p < 0.001) and male (fat: r = 0.91; muscle: r = 0.71; p < 0.001) patients. Linear regression analysis was statistically significant (fat: R 2 = 0.73 (female) and 0.74 (male); muscle: R 2 = 0.56 (female) and 0.56 (male); p < 0.001), showing that BFI CT and LSMI CT allowed prediction of BFP BIA and SMI BIA for both sexes.
Conclusion
CT-based BCA strongly correlates with BIA results and yields quantitative results for BFP and SMI comparable to the existing gold standard. Key Points Question CT-based body composition analysis (BCA) is moving more and more into clinical focus, but validation against established methods is lacking.
Findings
Fully automated CT-based BCA correlates very strongly with guideline-accepted bioelectrical impedance analysis (BIA). Clinical relevance BCA is currently moving further into clinical focus to improve assessment of patient frailty and individualize therapies accordingly. Comparability with established BIA strengthens the value of CT-based BCA and supports its translation into clinical routine. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Radiology, Berlin, Germany. uli.fehrenbach@charite.de.
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Radiology, Berlin, Germany.
- Berlin Institute of Health at Charité-Universitätsmedizin Berlin, BIH Biomedical Innovation Academy, Berlin, Germany.
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Department of Hepatology & Gastroenterology, Campus Virchow Klinikum and Campus Charité Mitte, Berlin,…
Zitieren
APA
Fehrenbach, U., Hosse, C., Wienbrandt, W., Walter-Rittel, T., Kolck, J., Auer, T. A., Blüthner, E., Tacke, F., Beetz, N. L., & Geisel, D. (2025). Concordance between single-slice abdominal computed tomography-based and bioelectrical impedance-based analysis of body composition in a prospective study. European Radiology, 35(12), 8000-8011. https://doi.org/10.1007/s00330-025-11746-3
Vancouver
Fehrenbach U, Hosse C, Wienbrandt W, Walter-Rittel T, Kolck J, Auer TA, et al. Concordance between single-slice abdominal computed tomography-based and bioelectrical impedance-based analysis of body composition in a prospective study. European Radiology. 2025;35(12):8000-8011. doi:10.1007/s00330-025-11746-3
BibTeX
@article{Fehrenbach2025,
author = {Fehrenbach, Uli and Hosse, Clarissa and Wienbrandt, William and Walter-Rittel, Thula and Kolck, Johannes and Auer, Timo Alexander and Blüthner, Elisabeth and Tacke, Frank and Beetz, Nick Lasse and Geisel, Dominik},
title = {Concordance between single-slice abdominal computed tomography-based and bioelectrical impedance-based analysis of body composition in a prospective study},
journal = {European Radiology},
year = {2025},
volume = {35},
number = {12},
pages = {8000--8011},
doi = {10.1007/s00330-025-11746-3}
}
EUVIMED