An unenhanced 3D-FLAIR sequence using long repetition time and constant flip angle to image endolymphatic hydrops

Menglong Zhao, Huaili Jiang, Shujie Zhang, Kai Liu, Lei Zhou, Di Wu, Xixi Wen, Junpu Hu, Xuan Wang, Zhuang Liu, Yan Sha, Mengsu Zeng

European Radiology · 10.07.2024 · 35 (1) · 29-37

DOI: 10.1007/s00330-024-10906-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Radiology, online erschienen am 10.07.2024. Verfasst von Menglong Zhao und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Zhongshan Hospital, Fudan University, and Shanghai Institute of Medical Imaging, Shanghai, 200032, China. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Endolymphatic Hydrops, Phantoms, Imaging, Magnetic Resonance Imaging. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Shanghai Municipal Health Planning Commission Youth Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: To evaluate a three-dimensional fluid-attenuated inversion recovery (3D-FLAIR) sequence using a long repetition time (TR) and constant flip angle (CFA) in differentiating between perilymph and endolymph in a phantom study, and unenhanced endolymphatic hydrops (EH) imaging in a patient study. METHODS: Three solutions in similar ion and protein concentrations with endolymph, perilymph, and cerebrospinal fluid were prepared for variable flip angle (VFA) 3D-FLAIR (TR 10,000 ms) and CFA (120°) 3D-FLAIR using different TR (10,000, 16,000, and 20,000 ms). Fifty-two patients with probable or definite Meniere's disease received unenhanced CFA (120°) 3D-FLAIR using a long TR (20,000 ms) and 4-h-delay enhanced CFA (120°) 3D-FLAIR (TR 16,000 ms). Image quality, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of them were compared. Agreement in the evaluation of the EH degree between them was analyzed. RESULTS: In the phantom study, CNRs between perilymphatic and endolymphatic samples of VFA 3D-FLAIR (TR 10,000 ms) and CFA 3D-FLAIR (TR 10,000, 16,000, and 20,000 ms) were 6.66 ± 1.30, 17.90 ± 2.76, 23.87 ± 3.09, and 28.22 ± 3.15 (p < 0.001). In patient study, average score (3.65 ± 0.48 vs. 4.19 ± 0.40), SNR (34.56 ± 9.80 vs. 51.40 ± 11.27), and CNR (30.66 ± 10.55 vs. 45.08 ± 12.27) of unenhanced 3D-FLAIR were lower than enhanced 3D-FLAIR (p < 0.001). Evaluations of the two sequences showed excellent agreement in the cochlear and vestibule (Kappa value: 0.898 and 0.909). CONCLUSIONS: The CFA 3D-FLAIR sequence using a long TR could be used in unenhanced EH imaging with high accuracy. CLINICAL RELEVANCE STATEMENT: Unenhanced imaging of endolymphatic hydrops is valuable in the diagnosis and follow-up of patients, especially those who cannot receive contrast-enhanced MRI. KEY POINTS: Ion and protein concentration differences can be utilized in differentiating endolymph and perilymph on MRI. Endolymphatic and perilymphatic samples could be differentiated in vitro on this 3D-FLAIR sequence. This unenhanced 3D-FLAIR sequence is in excellent agreement with the enhanced constant flip angle 3D-FLAIR sequence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalEndolymphatic HydropsPhantoms, ImagingMagnetic Resonance ImagingSignal-To-Noise RatioHumansMaleFemaleMiddle AgedAdultAgedMeniere DiseasePerilymphReproducibility of Results

Stichwörter der Autor:innen

Ear (inner)Endolymphatic hydropsGadoliniumMagnetic resonance imaging

Autor:innen und Institutionen

  • Department of Radiology, Zhongshan Hospital, Fudan University, and Shanghai Institute of Medical Imaging, Shanghai, 200032, China.
  • Otolaryngology Head and Neck Surgery, Zhongshan Hospital, Fudan University, Shanghai, 200032, China.
  • United Imaging Healthcare, Shanghai, 201807, China.
  • Department of Radiology, Fudan University Shanghai Cancer Center, Department of Oncology, Shanghai Medical College, Fudan University, Shanghai, 200032, China. liuzhuang201010@126.com.
  • Department of Radiology, Eye and ENT Hospital, Fudan University, Shanghai, 200031, China. cjr.shayan@vip.163.com.
  • Department of Radiology, Zhongshan Hospital, Fudan University, and Shanghai Institute of Medical Imaging, Shanghai, 200032, China. zeng_mengsu@163.com.

Förderung

  • National Natural Science Foundation of China
  • Shanghai Municipal Health Planning Commission Youth Project

Zitieren

APA

Zhao, M., Jiang, H., Zhang, S., Liu, K., Zhou, L., Wu, D., Wen, X., Hu, J., Wang, X., Liu, Z., Sha, Y., & Zeng, M. (2024). An unenhanced 3D-FLAIR sequence using long repetition time and constant flip angle to image endolymphatic hydrops. European Radiology, 35(1), 29-37. https://doi.org/10.1007/s00330-024-10906-1

Vancouver

Zhao M, Jiang H, Zhang S, Liu K, Zhou L, Wu D, et al. An unenhanced 3D-FLAIR sequence using long repetition time and constant flip angle to image endolymphatic hydrops. European Radiology. 2024;35(1):29-37. doi:10.1007/s00330-024-10906-1

BibTeX
@article{Zhao2024,
  author = {Zhao, Menglong and Jiang, Huaili and Zhang, Shujie and Liu, Kai and Zhou, Lei and Wu, Di and Wen, Xixi and Hu, Junpu and Wang, Xuan and Liu, Zhuang and Sha, Yan and Zeng, Mengsu},
  title = {An unenhanced 3D-FLAIR sequence using long repetition time and constant flip angle to image endolymphatic hydrops},
  journal = {European Radiology},
  year = {2024},
  volume = {35},
  number = {1},
  pages = {29--37},
  doi = {10.1007/s00330-024-10906-1}
}
Kostenlos starten →