Improvement of coronary stent visualization using ultra-high-resolution photon-counting detector CT

Le Qin, Shanshui Zhou, Haipeng Dong, Jiqiang Li, Ruiyan Zhang, Chendie Yang, Peng Liu, Zhihan Xu, Fuhua Yan, Wenjie Yang

European Radiology · 27.04.2024 · 34 (10) · 6568-6577

DOI: 10.1007/s00330-024-10760-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Radiology, online erschienen am 27.04.2024. Verfasst von Le Qin und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China. Hauptthemen laut MEDLINE: Stents, Computed Tomography Angiography, Coronary Angiography. Die Arbeit wurde bisher 51-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: Natural Science Foundation of China under Grant.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: This study aimed to compare the image quality and diagnostic performance of standard-resolution (SR) and ultra-high-resolution (UHR) coronary CT angiography (CCTA) based on photon-counting detector CT (PCD-CT) of coronary stents and explore the best reconstruction kernel for stent imaging. METHODS: From July 2023 to September 2023, patients were enrolled to undergo CCTA using a dual-source PCD-CT system after coronary angioplasty with stent placement. SR images with a slice thickness/increment of 0.6/0.4 mm were reconstructed using a vascular kernel (Bv48), while UHR images with a slice thickness/increment of 0.2/0.2 mm were reconstructed using vascular kernels of six sharpness levels (Bv48, Bv56, Bv60, Bv64, Bv72, and Bv76). The in-stent lumen diameters were evaluated. Subjective image quality was also evaluated by a 5-point Likert scale. Invasive coronary angiography was conducted in 12 patients (25 stents). RESULTS: Sixty-nine patients (68.0 [61.0, 73.0] years, 46 males) with 131 stents were included. All UHR images had significantly larger in-stent lumen diameter than SR images (p < 0.001). Specifically, UHR-Bv72 and UHR-Bv76 for in-stent lumen diameter (2.17 [1.93, 2.63] mm versus 2.20 [1.93, 2.59] mm) ranked the two best kernels. The subjective analysis demonstrated that UHR-Bv72 images had the most pronounced effect on reducing blooming artifacts, showcasing in-stent lumen and stent demonstration, and diagnostic confidence (p < 0.001). Furthermore, SR and UHR-Bv72 images showed a diagnostic accuracy of 78.3% (95% confidence interval [CI]: 56.3%-92.5%) and 88.0% (95%CI: 68.8%-97.5%), respectively. CONCLUSION: UHR CCTA by PCD-CT leads to significantly improved visualization and diagnostic performance of coronary stents, and Bv72 is the optimal reconstruction kernel showing the stent struts and in-stent lumen. CLINICAL RELEVANCE STATEMENT: The significantly improved visualization of coronary stents using ultra-high resolution CCTA could increase the diagnostic accuracy for in-stent restenosis and avoid unnecessary invasive quantitative coronary angiography, thus changing the clinical management for patients after percutaneous coronary intervention. KEY POINTS: Coronary stent imaging is challenging with energy-integrating detector CT due to "blooming artifacts." UHR images using a PCD-CT enhanced coronary stent visualization. UHR coronary stent imaging demonstrated improved diagnostic accuracy in clinical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

StentsComputed Tomography AngiographyCoronary AngiographyHumansFemaleMaleAgedMiddle AgedPhotonsCoronary Artery Disease

Stichwörter der Autor:innen

Computed tomography angiographyCoronary diseasePercutaneous coronary interventionStentsTomography (X-ray computed)

Autor:innen und Institutionen

  • Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • College of Health Science and Technology, Shanghai Jiao Tong University School of Medicine, No. 150 Ruijin Er Road, Shanghai, 200025, China.
  • Department of Cardiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • CT Collaboration, Siemens Healthineers, 399 West Haiyang Road, Shanghai, 200126, China.
  • Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China. lisa_ywj@163.com.

Förderung

  • Natural Science Foundation of China under Grant

Zitieren

APA

Qin, L., Zhou, S., Dong, H., Li, J., Zhang, R., Yang, C., Liu, P., Xu, Z., Yan, F., & Yang, W. (2024). Improvement of coronary stent visualization using ultra-high-resolution photon-counting detector CT. European Radiology, 34(10), 6568-6577. https://doi.org/10.1007/s00330-024-10760-1

Vancouver

Qin L, Zhou S, Dong H, Li J, Zhang R, Yang C, et al. Improvement of coronary stent visualization using ultra-high-resolution photon-counting detector CT. European Radiology. 2024;34(10):6568-6577. doi:10.1007/s00330-024-10760-1

BibTeX
@article{Qin2024,
  author = {Qin, Le and Zhou, Shanshui and Dong, Haipeng and Li, Jiqiang and Zhang, Ruiyan and Yang, Chendie and Liu, Peng and Xu, Zhihan and Yan, Fuhua and Yang, Wenjie},
  title = {Improvement of coronary stent visualization using ultra-high-resolution photon-counting detector CT},
  journal = {European Radiology},
  year = {2024},
  volume = {34},
  number = {10},
  pages = {6568--6577},
  doi = {10.1007/s00330-024-10760-1}
}
Kostenlos starten →