EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Quantitative Assessment of Aortic Arch Conformability and Clinical Outcomes Following Thoracic Endovascular Aortic Repair

Sara Ffrench-Constant, Neda Oskooee, Safa Salim, Celia Riga, Richard Gibbs, Colin Bicknell, Michael Jenkins, Mohamad Hamady

CardioVascular and Interventional Radiology · 27.05.2026 · 49 (8) · 1488-1497

DOI: 10.1007/s00270-026-04481-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift CardioVascular and Interventional Radiology, online erschienen am 27.05.2026. Verfasst von Sara Ffrench-Constant und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Interventional Radiology, Imperial College Healthcare NHS Trust, London, UK. Hauptthemen laut MEDLINE: Aorta, Thoracic, Stents, Endovascular Aneurysm Repair, Blood Vessel Prosthesis. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Thoracic stent graft conformability is believed to provide mechanical advantages during thoracic endovascular aortic repair (TEVAR); however, quantitative evidence and associated clinical outcomes remain limited. This study aimed to assess stent graft conformability across varying aortic arch configurations and pathologies, and to evaluate short- and mid-term technical and clinical outcomes. Materials and Methods A retrospective, single-centre review was conducted of TEVAR procedures performed using the GORE cTAG stent graft over a seven-year period. Pre- and post-deployment anatomical measurements were obtained using Endosize® software to assess conformability across a range of aortic arch morphologies and aortic syndromes. Conformability was defined as a < 10% change in predefined anatomical parameters, including aortic arch angle and proximal landing zone (PLZ) angle. Demographic data, mortality, complications, incidence of bird beaking, and reintervention rates were analysed.

Results

Of 189 TEVAR procedures performed during the study period, 111 met inclusion criteria and utilised the GORE cTAG stent graft. High conformability was observed following intervention, with 79% of patients demonstrating < 10% change in aortic arch angle and 88% demonstrating < 10% change in PLZ angle. Bird beaking occurred in 10.8% of patients. The overall reintervention rate was 5.4%, with greater angular changes significantly associated with reintervention ( p = 0.039). Overall survival rates at 30 days, 1 year, and 5 years were 95%, 86%, and 80%, respectively.

Conclusion

The GORE cTAG stent graft demonstrates high conformability across diverse aortic arch morphologies and pathologies, which is associated with favourable short- and mid-term technical and clinical outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Aorta, ThoracicStentsEndovascular Aneurysm RepairBlood Vessel ProsthesisEndovascular ProceduresAortic Aneurysm, ThoracicHumansRetrospective StudiesFemaleTreatment OutcomeMaleAgedMiddle AgedComputed Tomography AngiographyProsthesis DesignBlood Vessel Prosthesis ImplantationPostoperative Complications

Stichwörter der Autor:innen

AortaAortic arch angulationBirdbeakConformabilityDissectionDurabilityEndograftTEVARThoracic aorta

Autor:innen und Institutionen

  • Department of Interventional Radiology, Imperial College Healthcare NHS Trust, London, UK.
  • Department of Vascular Surgery, Imperial College Healthcare NHS Trust, London, UK.
  • Department of Surgery and Cancer, Imperial College London, London, UK.
  • Department of Interventional Radiology, Imperial College Healthcare NHS Trust, London, UK. m.hamady@imperial.ac.uk.
  • Department of Surgery and Cancer, Imperial College London, London, UK. m.hamady@imperial.ac.uk.

Zitieren

APA

Ffrench-Constant, S., Oskooee, N., Salim, S., Riga, C., Gibbs, R., Bicknell, C., Jenkins, M., & Hamady, M. (2026). Quantitative Assessment of Aortic Arch Conformability and Clinical Outcomes Following Thoracic Endovascular Aortic Repair. CardioVascular and Interventional Radiology, 49(8), 1488-1497. https://doi.org/10.1007/s00270-026-04481-9

Vancouver

Ffrench-Constant S, Oskooee N, Salim S, Riga C, Gibbs R, Bicknell C, et al. Quantitative Assessment of Aortic Arch Conformability and Clinical Outcomes Following Thoracic Endovascular Aortic Repair. CardioVascular and Interventional Radiology. 2026;49(8):1488-1497. doi:10.1007/s00270-026-04481-9

BibTeX
@article{FfrenchConstant2026,
  author = {Ffrench-Constant, Sara and Oskooee, Neda and Salim, Safa and Riga, Celia and Gibbs, Richard and Bicknell, Colin and Jenkins, Michael and Hamady, Mohamad},
  title = {Quantitative Assessment of Aortic Arch Conformability and Clinical Outcomes Following Thoracic Endovascular Aortic Repair},
  journal = {CardioVascular and Interventional Radiology},
  year = {2026},
  volume = {49},
  number = {8},
  pages = {1488--1497},
  doi = {10.1007/s00270-026-04481-9}
}
Kostenlos starten →