EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Evaluation of Fully Teleoperated Robotic Endovascular Interventions with Haptic Feedback: The SENTANTE Endovascular Robotic System

Emiel W M Huistra, Giovanni Federico Torsello, Konstantinos Stavroulakis, Lorenzo Patrone, Arturas Mackevicius, Max Amor, Vilius Dambrauskas, Vaidas Labunskas, Leona Damalakiene, Clark J Zeebregts, Tomas Baltrunas

CardioVascular and Interventional Radiology · 2026

DOI: 10.1007/s00270-026-04375-w ↗

Kurz gefasst

Fachartikel in der Zeitschrift CardioVascular and Interventional Radiology, 2026. Verfasst von Emiel W M Huistra und 10 weiteren Autor:innen. Beteiligte Einrichtung: Division of Vascular Surgery, Department of Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. w.m.hu…. Hauptthemen laut MEDLINE: Endovascular Procedures, Robotics, Robotic Surgical Procedures. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EIC Accelerator Blended Finance.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate a fully robotic, teleoperated system for performing various endovascular interventions in a preclinical porcine model. Materials and Methods The SENTANTE™ Endovascular Robotic System was assessed in six healthy porcine subjects using standard endovascular devices. Outcomes were assessed via procedural monitoring and gross pathological analysis. Haptic feedback performance was qualitatively assessed by the operators using a four-point scale (1 = very good, 2 = fair, 3 = poor, 4 = inadequate).

Results

A total of 18 endovascular interventional procedures were successfully completed, including renal artery stenting (3), embolization of the superior right renal artery branch (3), percutaneous transluminal angioplasty (PTA) of the superior mesenteric artery (3), left vertebral artery stenting (3), contralateral iliac artery PTA (3), and ipsilateral iliac artery stenting (3). All procedures (100%) were completed without manual conversion, including pre and post-procedural angiography, resulting in near-zero operator radiation exposure. The system was compatible with standard guidewires (0.014″, 0.018″, and 0.035″), catheters (2.7–8 Fr), and pushable coils (0.018″ and 0.035″). Post-procedural angiography demonstrated no signs of vessel injury. Minimal vessel trauma was observed in 3 of 16 target vessels during gross necropsy. Haptic feedback was rated as very good in three test models and as fair in the remaining three.

Conclusions

In a preclinical model, the SENTANTE™ system demonstrated the feasibility of performing endovascular interventions via teleoperation, including angiography, stenting, coil embolization, and PTA. Its compatibility with standard devices, haptic feedback, and elimination of operator radiation exposure support progression to clinical feasibility trials. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endovascular ProceduresRoboticsRobotic Surgical ProceduresAnimalsSwineStentsEquipment DesignEmbolization, TherapeuticFemaleTelemedicineFeedback

Stichwörter der Autor:innen

EndovascularInterventional radiologyRobotic surgical proceduresRoboticsTeleoperation

Autor:innen und Institutionen

  • Division of Vascular Surgery, Department of Surgery, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. w.m.huistra@umcg.nl.
  • Department of Clinical and Interventional Radiology, Göttingen University Medical Center, Göttingen, Germany.
  • Department of Vascular Surgery, Ludwig Maximilians University, Munich, Germany.
  • Vascular and Endovascular Surgery, Mathias Spital Rheine, Rheine, Germany.
  • Vascular and Endovascular Surgery Unit, San Giovanni Di Dio Hospital, Florence, Italy.
  • Department of Vascular Surgery, Republican Vilnius University Hospital, Vilnius, Lithuania.

Förderung

  • HORIZON EIC Accelerator Blended Finance

Zitieren

APA

Huistra, E. W. M., Torsello, G. F., Stavroulakis, K., Patrone, L., Mackevicius, A., Amor, M., Dambrauskas, V., Labunskas, V., Damalakiene, L., Zeebregts, C. J., & Baltrunas, T. (2026). Evaluation of Fully Teleoperated Robotic Endovascular Interventions with Haptic Feedback: The SENTANTE Endovascular Robotic System. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-026-04375-w

Vancouver

Huistra EWM, Torsello GF, Stavroulakis K, Patrone L, Mackevicius A, Amor M, et al. Evaluation of Fully Teleoperated Robotic Endovascular Interventions with Haptic Feedback: The SENTANTE Endovascular Robotic System. CardioVascular and Interventional Radiology. 2026. doi:10.1007/s00270-026-04375-w

BibTeX
@article{Huistra2026,
  author = {Huistra, Emiel W M and Torsello, Giovanni Federico and Stavroulakis, Konstantinos and Patrone, Lorenzo and Mackevicius, Arturas and Amor, Max and Dambrauskas, Vilius and Labunskas, Vaidas and Damalakiene, Leona and Zeebregts, Clark J and Baltrunas, Tomas},
  title = {Evaluation of Fully Teleoperated Robotic Endovascular Interventions with Haptic Feedback: The SENTANTE Endovascular Robotic System},
  journal = {CardioVascular and Interventional Radiology},
  year = {2026},
  doi = {10.1007/s00270-026-04375-w}
}
Kostenlos starten →