Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo Predictability Assessment

Marco Calandri, Susan Van Der Lei, Valentina Cignini, Timothy Neuss, Alessandro Serafini, Gaetano Risi, Francesca Menchini, Fernanda Ciferri, Marco Gatti, Carlo Gazzera, Paolo Fonio, Roberta Sirovich und 1 weitere

CardioVascular and Interventional Radiology · 2026

DOI: 10.1007/s00270-025-04342-x ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift CardioVascular and Interventional Radiology, 2026. Verfasst von Marco Calandri und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgical Sciences, University of Turin, Via Genova 3, 10126, Turin, Italy. Hauptthemen laut MEDLINE: Liver Neoplasms, Microwaves, Catheter Ablation, Surgery, Computer-Assisted. Die Arbeit nennt 22 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To assess the predictability of Emprint HP-Generator (150 W, 2.45 GHz) in terms of ablation volume and sphericity both in the ex vivo and in vivo setting. MATERIALS AND METHODS: In an ex vivo setting, 48 ablations were performed on porcine livers at varying ablation times with a constant power of 150 W. MRI assessment of the ablation volumes was performed. In the in vivo setting consecutive procedures with 150W application with no rampage, single antenna positioning and availability of a contrast-enhanced computed tomography (CECT) immediately after the ablation were included. A semi-automated segmentation tool was used for volume estimation and spherical index (SI) was calculated. RESULTS: In the ex vivo group, mean ablation time was 5.3 ± 2.5 min (range 1-10 min), mean ablation diameter was 37.7 ± 5.8 mm (range 21.3-49 mm) and mean ablation volume was 20.1 ± 10.8 cm3 (range 3-38.4 cm3) with a mean SI of 0.83 ± 0.08. In the in vivo group 78 percutaneous liver ablations (tumour mean diameter 23.2 ± 11 mm) in 65 patients (M46; F19; mean age 66.8 ± 11.3) were retrospectively enrolled from two institutions. The mean ablation time was 4,9 ± 2,5 min (range 1-10 min), mean ablation diameter was 41.3 ± 9.6 mm (range 16-73 mm) and mean ablation volume was 27.8 ± 15.6 cm3(range 6.4-105.6 cm3). Mean SI was 0.85 ± 0.11. Ablation volumes tend to be larger in case of hepatocellular carcinoma (HCC) compared to other histology (33.7 ± 21.2 cm3 vs 24.8 ± 10.5 cm3; p = 0.08). Both in ex vivo and in vivo settings, the SI remained stable regardless of ablation time and volume (Pearson's r 0.08, p 0.46), while ablation volumes increased linearly with time (Pearson's r 0.79, p < 0.001). Clavien-Dindo 1 complications were registered (4/65, 6.1%). CONCLUSION: The Emprint HP-generator enables the creation of large predictable volumes in both settings. SI remains consistent regardless of ablation time and volume.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Liver NeoplasmsMicrowavesCatheter AblationSurgery, Computer-AssistedAnimalsSwineTomography, X-Ray ComputedContrast MediaLiverMagnetic Resonance Imaging

Stichwörter der Autor:innen

Colorectal liver metastasesHepatocellular carcinomaInterventional OncologyLiver tumoursMicrowave ablation (MWA)SphericityThermal ablationThermosphere technology

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Surgical Sciences, University of Turin, Via Genova 3, 10126, Turin, Italy.
  • Department of Diagnostic and Interventional Radiology, A.O.U. Città Della Salute E Della Scienza Di Torino, Corso Bramante 88, 10126, Turin, Italy.
  • Department of Radiology and Nucleair Medicine, Amsterdam UMC Location Vrije Universiteit Amsterdam, Boelelaan 1117, Amsterdam, The Netherlands.
  • Cancer Centre Amsterdam, Treatment and Quality of Life, Amsterdam, The Netherlands.
  • Vascular and Interventional Radiology Unit, ASL Viterbo - Santa Rosa Hospital, Str. Sammartinese, 01100, Viterbo, Italy. valentinacignini@gmail.com.
  • Diagnostic and Interventional Radiology Unit, IRCCS Candiolo Institute, SP142 Km 3,95Candiolo, 10060, Turin, TO, Italy.

Zitieren

APA

Calandri, M., Van Der Lei, S., Cignini, V., Neuss, T., Serafini, A., Risi, G., Menchini, F., Ciferri, F., Gatti, M., Gazzera, C., Fonio, P., Sirovich, R., & Meijerink, M. R. (2026). Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo Predictability Assessment. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-025-04342-x

Vancouver

Calandri M, Van Der Lei S, Cignini V, Neuss T, Serafini A, Risi G, et al. Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo Predictability Assessment. CardioVascular and Interventional Radiology. 2026. doi:10.1007/s00270-025-04342-x

BibTeX
@article{Calandri2026,
  author = {Calandri, Marco and Van Der Lei, Susan and Cignini, Valentina and Neuss, Timothy and Serafini, Alessandro and Risi, Gaetano and Menchini, Francesca and Ciferri, Fernanda and Gatti, Marco and Gazzera, Carlo and Fonio, Paolo and Sirovich, Roberta and Meijerink, Martijn Ruben},
  title = {Image-Guided Microwave Ablation of Liver Tumours Using a 150-W Generator: Ex Vivo and Multicenter in Vivo Predictability Assessment},
  journal = {CardioVascular and Interventional Radiology},
  year = {2026},
  doi = {10.1007/s00270-025-04342-x}
}
Kostenlos starten →