Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model
CardioVascular and Interventional Radiology · 2023
DOI: 10.1007/s00270-023-03544-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift CardioVascular and Interventional Radiology, 2023. Verfasst von PengXu Ding und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Intervention, First Affiliated Hospital of Zhengzhou University, Zhengzhou, China. Hauptthemen laut MEDLINE: Portasystemic Shunt, Transjugular Intrahepatic. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: C. R. Bard, a wholly owned subsidiary of Becton, Dickinson and Company.
Abstract (Originalsprache)
Purpose
This study aimed to assess the safety, effectiveness, and feasibility of the Liverty™ transjugular intrahepatic portosystemic shunt (TIPS) access set, which has an ergonomic handle that allows for in situ cannula tip deflection and a distal steerable cannula angle, versus the COOK® Rosch-Uchida Transjugular Liver Access Set (RUPS-100) in healthy pigs.
Methods
Twelve pigs randomly underwent TIPS with the Liverty™ set or the RUPS-100 set. Three interventionalists performed 4 TIPS procedures, 2 with each set. The primary outcome was procedural success, defined as successful establishment of the intrahepatic portosystemic shunt and stent placement.
Results
The shunt was successfully established in 11 pigs. The procedural success was achieved in all 6 pigs in the Liverty™ group and 5 out of 6 pigs for the RUPS-100 group (Fisher exact test, P > 0.999). The mean duration of puncture was shorter in the Liverty™ group versus the RUPS-100 group (12.3 ± 4.5 min vs. 16.2 ± 8.5 min), but without significant statistical difference (two sample t test, P = 0.359). The cannula angle was adjusted 69% of passes in the Liverty™ group, which was significantly higher than that in the RUPS-100 group (12%, P = 0.004). Overall, the TIPS procedural performance was comparable between the groups. Both sets were safe. No intraabdominal hemorrhage, vascular injuries, tissue or organ injuries, porto-biliary fistula, biliary peritonitis, and infection or abscess occurred in either group.
Conclusion
The Liverty™ set is safe and has similar procedural metrics to the COOK® RUPS-100 set. It allows in situ adjustment of the angle of the stiffening cannula without increasing procedure time and lessens the occurrences of periprocedural complications. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Intervention, First Affiliated Hospital of Zhengzhou University, Zhengzhou, China.
- Department of Radiology, Shengjing Hospital, China Medical University, Shenyang, China.
- Medical Affairs, Becton, Dickinson and Company, Shanghai, China.
- Research & Development, Becton, Dickinson and Company, Shanghai, China.
- Animal Lab, Shanghai Harborside Medical Technology Co.,Ltd, Shanghai, China.
- Department of Gastroenterology, Nanjing Drum Tower Hospital, The Affiliated Hospital of Nanjing University Medical School, 321 Zhongshan Road, Gulou District, Nanjing, 210008, China. josedy@126.com.
Förderung
- C. R. Bard, a wholly owned subsidiary of Becton, Dickinson and Company
Zitieren
APA
Ding, P., Ma, Y., Zhu, X., Wu, Y., Ong, J., Liu, P., Xiao, J., & Zhuge, Y. (2023). Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-023-03544-5
Vancouver
Ding P, Ma Y, Zhu X, Wu Y, Ong J, Liu P, et al. Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model. CardioVascular and Interventional Radiology. 2023. doi:10.1007/s00270-023-03544-5
BibTeX
@article{Ding2023,
author = {Ding, PengXu and Ma, Yujia and Zhu, Xiaoxia and Wu, Yijie and Ong, John and Liu, Pu and Xiao, Jiangqiang and Zhuge, Yuzheng},
title = {Safety and Effectiveness of a Novel Tips Access Set with Steerable Cannula in a Swine Model},
journal = {CardioVascular and Interventional Radiology},
year = {2023},
doi = {10.1007/s00270-023-03544-5}
}
EUVIMED