Repeatability of semi-quantitative [68Ga]Ga-FAPI-46 PET/CT measurements in pancreatobiliary cancers: a test-retest study
European Journal of Nuclear Medicine and Molecular Imaging · 2026-05 · 53 (6) · 3827-3837
DOI: 10.1007/s00259-025-07738-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Nuclear Medicine and Molecular Imaging, online erschienen am 16.01.2026. Verfasst von Rutger B Henrar und 12 weiteren Autor:innen. Beteiligte Einrichtung: Amsterdam UMC location Vrije Universiteit Amsterdam, De Boelelaan 1117, Amsterdam, The Netherlands. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Pancreatic Neoplasms, Biliary Tract Neoplasms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: KWF Kankerbestrijding, Cancer Center Amsterdam.
Abstract (Originalsprache)
Purpose
Radiolabelled fibroblast activation protein inhibitors (FAPI) are promising molecular imaging tracers for the diagnosis and staging of pancreatobiliary cancer. To enable treatment response assessment with [ 68 Ga]Ga-FAPI-46, the day-to-day variability of its uptake must be defined. This study aimed to determine the repeatability of semi-quantitative [ 68 Ga]Ga-FAPI-46 PET/CT measurements in pancreatobiliary cancers.
Methods
Patients with pathologically confirmed pancreatobiliary cancer (pancreatic ductal adenocarcinoma (PDAC) or cholangiocarcinoma (CCA)) were included. Patients underwent two [ 68 Ga]Ga-FAPI-46 PET/CT scans, within 7 days, without any treatment before or between scans. Static long axial field-of-view (LAFOV) PET/CT scans were performed 60 min post-injection of [ 68 Ga]Ga-FAPI-46 (EARL2-compliant reconstruction). Suspected malignant lesions were semi-automatically delineated (local background-adjusted 50% isocontour of the peak standardised uptake value (SUV)) and semi-quantitative measurements were extracted (SUV mean , SUV peak and SUV max ). Tumour-to-background ratios (TBR) were also calculated. Repeatability was assessed using the repeatability coefficient (RC) and the intraclass correlation coefficient (ICC).
Results
Twelve patients were included (seven PDAC, three perihilar CCA and two intrahepatic CCA). The median injected dose was 216 MBq (interquartile range (IQR) 167–266 MBq) and the median uptake time was 60 min (range 60–68). In total, 70 FAPI-positive lesions were delineated. The RCs of SUV mean , SUV peak and SUV max were 23.7%, 23.9% and 29.8%, respectively. For the blood pool adjusted TBR mean , TBR peak and TBR max , the RC’s were 21.6%, 20.8% and 28.0%, respectively. All ICCs were above 0.98.
Conclusion
Semi-quantitative measurements of [ 68 Ga]Ga-FAPI-46 PET/CT have an excellent repeatability and can potentially be used in future studies to assess treatment response in pancreatobiliary cancers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Amsterdam UMC location Vrije Universiteit Amsterdam, De Boelelaan 1117, Amsterdam, The Netherlands.
- Cancer Center Amsterdam, Imaging and Biomarkers, Amsterdam, The Netherlands.
- Univ Rennes, CLCC Eugène Marquis, INSERM, LTSI - UMR 1099, Rennes, France.
- Leiden University Medical Center, Albinusdreef 2, Leiden, the Netherlands.
- University of Twente, Drienerlolaan 5, Enschede, The Netherlands.
- Delft University of Technology, Mekelweg 6, Delft, The Netherlands.
ORCID-Profile: Rutger-Jan Swijnenburg
Förderung
- KWF Kankerbestrijding
- Cancer Center Amsterdam
Zitieren
APA
Henrar, R. B., Cysouw, M. C. F., Palard-Novello, X., Schwarte, L. A., Raijmakers, P. G. H. M., de Geus-Oei, L. F., Vahrmeijer, A. L., Kazemier, G., Windhorst, A. D., Boellaard, R., Yaqub, M., Oprea-Lager, D. E., & Swijnenburg, R. J. (2026). Repeatability of semi-quantitative [68Ga]Ga-FAPI-46 PET/CT measurements in pancreatobiliary cancers: a test-retest study. European Journal of Nuclear Medicine and Molecular Imaging, 53(6), 3827-3837. https://doi.org/10.1007/s00259-025-07738-6
Vancouver
Henrar RB, Cysouw MCF, Palard-Novello X, Schwarte LA, Raijmakers PGHM, de Geus-Oei LF, et al. Repeatability of semi-quantitative [68Ga]Ga-FAPI-46 PET/CT measurements in pancreatobiliary cancers: a test-retest study. European Journal of Nuclear Medicine and Molecular Imaging. 2026;53(6):3827-3837. doi:10.1007/s00259-025-07738-6
BibTeX
@article{Henrar2026,
author = {Henrar, Rutger B and Cysouw, Matthijs C F and Palard-Novello, Xavier and Schwarte, Lothar A and Raijmakers, Pieter G H M and de Geus-Oei, Lioe-Fee and Vahrmeijer, Alexander L and Kazemier, Geert and Windhorst, Albert D and Boellaard, Ronald and Yaqub, Maqsood and Oprea-Lager, Daniela E and Swijnenburg, Rutger-Jan},
title = {Repeatability of semi-quantitative [68Ga]Ga-FAPI-46 PET/CT measurements in pancreatobiliary cancers: a test-retest study},
journal = {European Journal of Nuclear Medicine and Molecular Imaging},
year = {2026},
volume = {53},
number = {6},
pages = {3827--3837},
doi = {10.1007/s00259-025-07738-6}
}