Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning
European Journal of Nuclear Medicine and Molecular Imaging · 2025
DOI: 10.1007/s00259-025-07086-5 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift European Journal of Nuclear Medicine and Molecular Imaging, 2025. Verfasst von Hanzhong Wang und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. Hauptthemen laut MEDLINE: Deep Learning, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Patient Positioning. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: National leading clinical discipline project and Shanghai Municipal Key Clinical Specialty.
Abstract (Originalsprache)
PURPOSE: To address the challenges of verifying MR-based attenuation correction (MRAC) in PET/MR due to CT positional mismatches and alignment issues, this study utilized a flatbed insert and arms-down positioning during PET/CT scans to achieve precise MR-CT matching for accurate MRAC evaluation. METHODS: A validation dataset of 21 patients underwent whole-body [18F]FDG PET/CT followed by [18F]FDG PET/MR. A flatbed insert ensured consistent positioning, allowing direct comparison of four MRAC methods-four-tissue and five-tissue models with discrete and continuous μ-maps-against CT-based attenuation correction (CTAC). A deep learning-based framework, trained on a dataset of 300 patients, was used to generate synthesized-CTs from MR images, forming the basis for all MRAC methods. Quantitative analyses were conducted at the whole-body, region of interest, and lesion levels, with lesion-distance analysis evaluating the impact of bone proximity on standardized uptake value (SUV) quantification. RESULTS: Distinct differences were observed among MRAC methods in spine and femur regions. Joint histogram analysis showed MRAC-4 (continuous μ-map) closely aligned with CTAC. Lesion-distance analysis revealed MRAC-4 minimized bone-induced SUV interference (r = 0.01, p = 0.8643). However, tissues prone to bone segmentation interference, such as the spine and liver, exhibited greater SUV variability and lower reproducibility in MRAC-4 compared to MRAC-2 (2D bone segmentation, discrete μ-map) and MRAC-3 (3D bone segmentation, discrete μ-map). CONCLUSION: Using a flatbed insert, this study validated MRAC with high precision. Continuous μ-value MRAC method (MRAC-4) demonstrated superior accuracy and minimized bone-related SUV errors but faced challenges in reproducibility, particularly in bone-rich regions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nuclear Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
- Institute for Medical Imaging Technology, Ruijin Hospital, Shanghai Jiao Tong University, Shanghai, China.
- Central Research Institute, United Imaging Healthcare Group Co., Ltd, Shanghai, China.
- Beijing United Imaging Research Institute of Intelligent Imaging, Beijing, China.
- Department of Nuclear Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. wj21061@rjh.com.cn.
- Department of Nuclear Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. lb10363@rjh.com.cn.
ORCID-Profile: Biao Li
Förderung
- National leading clinical discipline project and Shanghai Municipal Key Clinical Specialty
Zitieren
APA
Wang, H., Wang, Y., Xue, Q., Zhang, Y., Qiao, X., Lin, Z., Zheng, J., Zhang, Z., Yang, Y., Zhang, M., Huang, Q., Huang, Y., Cao, T., Wang, J., & Li, B. (2025). Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning. European Journal of Nuclear Medicine and Molecular Imaging. https://doi.org/10.1007/s00259-025-07086-5
Vancouver
Wang H, Wang Y, Xue Q, Zhang Y, Qiao X, Lin Z, et al. Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning. European Journal of Nuclear Medicine and Molecular Imaging. 2025. doi:10.1007/s00259-025-07086-5
BibTeX
@article{Wang2025,
author = {Wang, Hanzhong and Wang, Yue and Xue, Qiaoyi and Zhang, Yu and Qiao, Xiaoya and Lin, Zengping and Zheng, Jiaxu and Zhang, Zheng and Yang, Yang and Zhang, Min and Huang, Qiu and Huang, Yanqi and Cao, Tuoyu and Wang, Jin and Li, Biao},
title = {Optimizing MR-based attenuation correction in hybrid PET/MR using deep learning: validation with a flatbed insert and consistent patient positioning},
journal = {European Journal of Nuclear Medicine and Molecular Imaging},
year = {2025},
doi = {10.1007/s00259-025-07086-5}
}
EUVIMED