Trabecular bone analysis: ultra-high-resolution CT goes far beyond high-resolution CT and gets closer to micro-CT (a study using Canon Medical CT devices)
Skeletal Radiology · 2026-01 · 55 (1) · 71-80
DOI: 10.1007/s00256-025-05001-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift Skeletal Radiology, online erschienen am 30.07.2025. Verfasst von Romain Gillet und 9 weiteren Autor:innen. Beteiligte Einrichtung: Guilloz Imaging Department, Central Hospital, University Hospital Center of Nancy, 54000, Nancy, France. r.gillet@chru-nancy.fr. Hauptthemen laut MEDLINE: Cancellous Bone, Tomography, X-Ray Computed, X-Ray Microtomography, Tibia. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE: High-resolution CT (HR-CT) cannot image trabecular bone due to insufficient spatial resolution. Ultra-high-resolution CT may be a valuable alternative. We aimed to describe the accuracy of Canon Medical HR, super-high-resolution (SHR), and ultra-high-resolution (UHR)-CT in measuring trabecular bone microarchitectural parameters using micro-CT as a reference. MATERIAL AND METHODS: Sixteen cadaveric distal tibial epiphyses were enrolled in this pre-clinical study. Images were acquired with HR-CT (i.e., 0.5 mm slice thickness/5122 matrix) and SHR-CT (i.e., 0.25 mm slice thickness and 10242 matrix) with and without deep learning reconstruction (DLR) and UHR-CT (i.e., 0.25 mm slice thickness/20482 matrix) without DLR. Trabecular bone parameters were compared. RESULTS: Trabecular thickness was closest with UHR-CT but remained 1.37 times that of micro-CT (P < 0.001). With SHR-CT without and with DLR, it was 1.75 and 1.79 times that of micro-CT, respectively (P < 0.001), and 3.58 and 3.68 times that of micro-CT with HR-CT without and with DLR, respectively (P < 0.001). Trabecular separation was 0.7 times that of micro-CT with UHR-CT (P < 0.001), 0.93 and 0.94 times that of micro-CT with SHR-CT without and with DLR (P = 0.36 and 0.79, respectively), and 1.52 and 1.36 times that of micro-CT with HR-CT without and with DLR (P < 0.001). Bone volume/total volume was overestimated (i.e., 1.66 to 1.92 times that of micro-CT) by all techniques (P < 0.001). However, HR-CT values were superior to UHR-CT values (P = 0.03 and 0.01, without and with DLR, respectively). CONCLUSION: UHR and SHR-CT were the closest techniques to micro-CT and surpassed HR-CT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Guilloz Imaging Department, Central Hospital, University Hospital Center of Nancy, 54000, Nancy, France. r.gillet@chru-nancy.fr.
- Université de Lorraine, INSERM, IADI, 54000, Nancy, France. r.gillet@chru-nancy.fr.
- CIC, Innovation Technologique, Université de Lorraine, University Hospital Center of Nancy, 54000, Nancy, France. r.gillet@chru-nancy.fr.
- Guilloz Imaging Department, Central Hospital, University Hospital Center of Nancy, 54000, Nancy, France.
- Université de Lorraine, INSERM, IADI, 54000, Nancy, France.
- CIC, Innovation Technologique, Université de Lorraine, University Hospital Center of Nancy, 54000, Nancy, France.
Zitieren
APA
Gillet, R., Puel, U., Amer, A., Doyen, M., Boubaker, F., Assabah, B., Hossu, G., Gillet, P., Blum, A., & Teixeira, P. A. G. (2026). Trabecular bone analysis: ultra-high-resolution CT goes far beyond high-resolution CT and gets closer to micro-CT (a study using Canon Medical CT devices). Skeletal Radiology, 55(1), 71-80. https://doi.org/10.1007/s00256-025-05001-5
Vancouver
Gillet R, Puel U, Amer A, Doyen M, Boubaker F, Assabah B, et al. Trabecular bone analysis: ultra-high-resolution CT goes far beyond high-resolution CT and gets closer to micro-CT (a study using Canon Medical CT devices). Skeletal Radiology. 2026;55(1):71-80. doi:10.1007/s00256-025-05001-5
BibTeX
@article{Gillet2026,
author = {Gillet, Romain and Puel, Ulysse and Amer, Anas and Doyen, Mathieu and Boubaker, Fatma and Assabah, Bouchra and Hossu, Gabriela and Gillet, Pierre and Blum, Alain and Teixeira, Pedro Augusto Gondim},
title = {Trabecular bone analysis: ultra-high-resolution CT goes far beyond high-resolution CT and gets closer to micro-CT (a study using Canon Medical CT devices)},
journal = {Skeletal Radiology},
year = {2026},
volume = {55},
number = {1},
pages = {71--80},
doi = {10.1007/s00256-025-05001-5}
}
EUVIMED