Reaction of KHP with excess NaOH or TRIS as standard reactions for calibration of titration calorimeters from 0 to 60 °C
European Biophysics Journal · 2024
DOI: 10.1007/s00249-024-01705-z ↗Kurz gefasst
Fachartikel in der Zeitschrift European Biophysics Journal, 2024. Verfasst von Jason D Kenealey und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nutrition, Dietetics and Food Science, Brigham Young University, Provo, UT, 84602, USA. Hauptthemen laut MEDLINE: Sodium Hydroxide, Calorimetry, Tromethamine. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Portuguese Foundation for Science and Technology.
Abstract (Originalsprache)
Calibration of titration calorimeters is an ongoing problem, particularly with calorimeters with reaction vessel volumes < 10 mL in which an electrical calibration heater is positioned outside the calorimetric vessel. Consequently, a chemical reaction with a known enthalpy change must be used to accurately calibrate these calorimeters. This work proposes the use of standard solutions of potassium acid phthalate (KHP) titrated into solutions of excess sodium hydroxide (NaOH) or excess tris(hydroxymethyl)aminomethane (TRIS) as standard reactions to determine the collective accuracy of the relevant variables in a determination of the molar enthalpy change for a reaction. KHP is readily available in high purity, weighable for easy preparation of solutions with accurately known concentrations, stable in solution, not compromised by side reactions with common contaminants such as atmospheric CO 2 , and non-corrosive to materials used in calorimeter construction. Molar enthalpy changes for these reactions were calculated from 0 to 60 °C from reliable literature data for the p K a of KHP, the molar enthalpy change for protonation of TRIS, and the molar enthalpy change for ionization of water. The feasibility of using these reactions as enthalpic standards was tested in several calorimeters; a 50 mL CSC 4300, a 185 μL NanoITC, a 1.4 mL VP-ITC, and a TAM III with 1 mL reaction vessels. The results from the 50 mL CSC 4300, which was accurately calibrated with an electric heater, verified the accuracy of the calculated standard values for the molar enthalpy changes of the proposed reactions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nutrition, Dietetics and Food Science, Brigham Young University, Provo, UT, 84602, USA.
- CIQUP, Institute of Molecular Sciences (IMS), Department of Chemistry and Biochemistry, Faculty of Sciences, University of Porto, Porto, Portugal.
- Operations Science and Technology, AbbVie, Inc., 1401 Sheridan Road, North Chicago, IL, NC-A460064, USA.
- Collaborative Innovation Center of Henan Province for Green Manufacturing of Fine Chemicals, Key Laboratory of Green Chemical Media and Reactions, Ministry of Education, School of Chemistry and Chemical Engineering, Hen…
- Department of Chemistry and Biochemistry, Brigham Young University, Provo, UT, 84602, USA. ldhansen@chem.byu.edu.
Förderung
- National Natural Science Foundation of China
- Portuguese Foundation for Science and Technology
Zitieren
APA
Kenealey, J. D., Bastos, M., Assaf, Z., Bai, G., Zhao, W., Jarrard, T., Tower, C., & Hansen, L. D. (2024). Reaction of KHP with excess NaOH or TRIS as standard reactions for calibration of titration calorimeters from 0 to 60 °C. European Biophysics Journal. https://doi.org/10.1007/s00249-024-01705-z
Vancouver
Kenealey JD, Bastos M, Assaf Z, Bai G, Zhao W, Jarrard T, et al. Reaction of KHP with excess NaOH or TRIS as standard reactions for calibration of titration calorimeters from 0 to 60 °C. European Biophysics Journal. 2024. doi:10.1007/s00249-024-01705-z
BibTeX
@article{Kenealey2024,
author = {Kenealey, Jason D and Bastos, Margarida and Assaf, Zaid and Bai, Guangyue and Zhao, Wenqi and Jarrard, Tyler and Tower, Colter and Hansen, Lee D},
title = {Reaction of KHP with excess NaOH or TRIS as standard reactions for calibration of titration calorimeters from 0 to 60 °C},
journal = {European Biophysics Journal},
year = {2024},
doi = {10.1007/s00249-024-01705-z}
}
EUVIMED