Impact of upgrading from a 25-cm to a 30-cm z-axis field of view digital PET/CT in a pediatric hospital
Pediatric Radiology · 2024
DOI: 10.1007/s00247-024-06049-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Radiology, 2024. Verfasst von Joseph G Meier und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Ave, MLC 5031, Cincinnati, OH, 45229, USA. joseph.meier@cchmc.org. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Hospitals, Pediatric, Radiopharmaceuticals. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Increased positron emission tomography (PET) scanner z -axis coverage provides an opportunity in pediatrics to reduce dose, anesthesia, or repeat scans due to motion.
Objective
Recently, our digital PET scanner was upgraded from a 25-cm to a 30-cm z -axis coverage. We compare the two systems through National Electrical Manufacturing Association (NEMA) testing and evaluation of paired images from patients scanned on both systems. Materials and methods NEMA testing and a retrospective review of pediatric patients who underwent clinically indicated 18F-fluorodeoxyglucose (FDG) PET computed tomography (PET/CT) on both systems with unchanged acquisition parameters were performed. Image quality was assessed with liver signal to noise ratio (SNR-liver) and contrast to noise ratio (CNR) in the thigh muscle and liver with results compared with an unpaired t -test. Three readers independently reviewed paired (25 cm and 30 cm) images from the same patient, blinded to scanner configuration.
Results
Expansion to 30 cm increased system sensitivity to 29.8% (23.4 cps/kBq to 30.4 cps/kBq). Seventeen patients (6 male/11 female, median age 12.5 (IQR 8.3–15.0) years, median weight 53.7 (IQR 34.2–68.7) kg) were included. SNR-liver and CNR increased by 35.1% (IQR 19.0–48.4%) and 43.1% (IQR 6.2–50.2%) ( P -value
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Ave, MLC 5031, Cincinnati, OH, 45229, USA. joseph.meier@cchmc.org.
- Department of Radiology, University of Cincinnati College of Medicine, 3230 Eden Ave, Cincinnati, OH, 45267, USA. joseph.meier@cchmc.org.
- Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Ave, MLC 5031, Cincinnati, OH, 45229, USA.
- Department of Radiology, University of Cincinnati College of Medicine, 3230 Eden Ave, Cincinnati, OH, 45267, USA.
- Department of Pediatrics, University of Cincinnati College of Medicine, 3230 Eden Ave, Cincinnati, OH, 45267, USA.
- Department of Radiology, AdventHealth Central Florida, Orlando, FL, USA.
ORCID-Profile: Joseph G Meier, Andrew T Trout, Elanchezhian Somasundaram, Samuel L Brady
Zitieren
APA
Meier, J. G., Trout, A. T., Abu Ata, N., Sharp, S. E., Anton, C. G., Somasundaram, E., & Brady, S. L. (2024). Impact of upgrading from a 25-cm to a 30-cm z-axis field of view digital PET/CT in a pediatric hospital. Pediatric Radiology. https://doi.org/10.1007/s00247-024-06049-6
Vancouver
Meier JG, Trout AT, Abu Ata N, Sharp SE, Anton CG, Somasundaram E, et al. Impact of upgrading from a 25-cm to a 30-cm z-axis field of view digital PET/CT in a pediatric hospital. Pediatric Radiology. 2024. doi:10.1007/s00247-024-06049-6
BibTeX
@article{Meier2024,
author = {Meier, Joseph G and Trout, Andrew T and Abu Ata, Nadeen and Sharp, Susan E and Anton, Christopher G and Somasundaram, Elanchezhian and Brady, Samuel L},
title = {Impact of upgrading from a 25-cm to a 30-cm z-axis field of view digital PET/CT in a pediatric hospital},
journal = {Pediatric Radiology},
year = {2024},
doi = {10.1007/s00247-024-06049-6}
}
EUVIMED