Epicardial Pacemaker Lead Related Cardiac Strangulation: The Importance of Early Recognition
Pediatric Cardiology · 2026
DOI: 10.1007/s00246-025-03792-x ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Cardiology, 2026. Verfasst von Scott Kendall und 8 weiteren Autor:innen. Beteiligte Einrichtung: Queens University Belfast, University Road, Belfast, UK. s.kendall@qub.ac.uk. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Electrodes, Implanted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 8 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Belfast Trust Charitable Funds, QUB.
Abstract (Originalsprache)
Lead strangulation is a dangerous complication of epicardial pacemaker insertion. This complication has been increasingly highlighted lately. Our institution has recently identified four cases over the past five years. This study’s aim was to 1) identify risk factors for strangulation and 2) prospectively screen existing epicardial pacemaker patients for unrecognized strangulation or features that would prompt closer review. Patients known to the pacemaker clinic with epicardial pacemakers inserted from 2005 to 2023 were included. Electronic health records were used to locate all subjects and gather data. Risk factors were identified using Firth’s penalized method of logistic regression. Forty-five patients were included, of which four (8.8%) had evidence of strangulation. Posterior–anterior (PA) chest radiographs all demonstrated characteristic looping patterns of the pacing leads, with confirmation on CT angiography. All affected patients underwent revision surgery. Implantation at a weight of less than 6.5 kg was associated with a significantly increased incidence of strangulation (OR 25, P 0.044). Other factors including lead length, presence of structural cardiac disease, and dual-chamber insertion were not statistically significant. No patients who were prospectively screened were found to have strangulation. Children undergoing insertion of a pacemaker early in infancy are at particularly high risk of strangulation and should be closely monitored following surgery. Regular chest radiography (every three years) to screen for this complication is advised. Larger multi-center studies to pool data for this relatively rare complication may help identify other risk factors for strangulation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Queens University Belfast, University Road, Belfast, UK. s.kendall@qub.ac.uk.
- Paediatric Cardiology Department, Royal Belfast Hospital for Sick Children, Belfast, UK.
- Queens University Belfast, University Road, Belfast, UK.
- Cardiothoracic Department, Children's Health Ireland at Crumlin, Dublin, Ireland.
Förderung
- Belfast Trust Charitable Funds
- QUB
Zitieren
APA
Kendall, S., Agouba, R., Murray, J., Morrison, M. L., McCrossan, B., Grant, B., Sands, A., Casey, F., & Nolke, L. (2026). Epicardial Pacemaker Lead Related Cardiac Strangulation: The Importance of Early Recognition. Pediatric Cardiology. https://doi.org/10.1007/s00246-025-03792-x
Vancouver
Kendall S, Agouba R, Murray J, Morrison ML, McCrossan B, Grant B, et al. Epicardial Pacemaker Lead Related Cardiac Strangulation: The Importance of Early Recognition. Pediatric Cardiology. 2026. doi:10.1007/s00246-025-03792-x
BibTeX
@article{Kendall2026,
author = {Kendall, Scott and Agouba, Rihab and Murray, Jane and Morrison, Margaret Louise and McCrossan, Brian and Grant, Brian and Sands, Andrew and Casey, Frank and Nolke, Lars},
title = {Epicardial Pacemaker Lead Related Cardiac Strangulation: The Importance of Early Recognition},
journal = {Pediatric Cardiology},
year = {2026},
doi = {10.1007/s00246-025-03792-x}
}
EUVIMED