Slippery when wet: The effect of a novel zwitterionic coating on silicone ureteral stents
Urolithiasis · 2026
DOI: 10.1007/s00240-026-02070-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Urolithiasis, 2026. Verfasst von Ye Jean Lim und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Loma Linda University Health, 11234 Anderson Street, Loma Linda, CA, 92354, USA. Hauptthemen laut MEDLINE: Stents, Silicones, Ureter, Coated Materials, Biocompatible. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Silicone ureteral stents provide anti-encrustation properties and patient comfort but are more difficult to insert and remove due to increased surface friction. Applying zwitterionic coatings to silicone stents may reduce friction and facilitate stent placement and removal. This study compared the frictional forces of zwitterionic-coated silicone stents with commercial stents using benchtop models. The static and kinetic frictional forces of five ureteral stents were measured: bare silicone stent without coating (BS), single outer-coated zwitterionic silicone stent (Z1), double-coated (inner and outer) zwitterionic silicone stent (Z2), Black Silicone Filiform stent (“Black Beauty”) (BB), and Universa Soft stent (US). Friction was assessed with a submerged guidewire model to isolate stent–guidewire friction and a porcine ureteral model to simulate intraoperative stent insertion and removal. Five trials per stent were performed using a force gauge integrated with an Arduino-controlled motor system. Statistical significance was assessed using linear mixed-effects models with stent type and trial as fixed effects and ureter specimen as a random intercept. Pairwise comparisons were adjusted using the Holm method. In the submerged guidewire model, Z2 had 99%, 47%, and 40% lower kinetic friction than BS, BB, and US, respectively. During stent insertion, Z2 had 71%, 43%, and 44% lower kinetic friction than BS, BB, and US, respectively. During stent removal, Z2 had 92%, 58%, and 60% lower kinetic friction than BS, BB, and US, respectively. Linear mixed-effects models demonstrated a significant effect of stent type for all six friction conditions (all p < 0.001). The double-coated (inner and outer) zwitterionic silicone stent (Z2) demonstrated the lowest frictional forces across all testing conditions. Zwitterionic surface modification may therefore reduce friction during silicone stent placement and removal.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Urology, Loma Linda University Health, 11234 Anderson Street, Loma Linda, CA, 92354, USA.
- Department of Urology, Loma Linda University Health, 11234 Anderson Street, Loma Linda, CA, 92354, USA. dbaldwin@llu.edu.
Zitieren
APA
Lim, Y. J., Lim, Y. N. D., Church, C. A., Haynal, E., Baldwin, E. A., & Baldwin, D. D. (2026). Slippery when wet: The effect of a novel zwitterionic coating on silicone ureteral stents. Urolithiasis. https://doi.org/10.1007/s00240-026-02070-6
Vancouver
Lim YJ, Lim YND, Church CA, Haynal E, Baldwin EA, Baldwin DD. Slippery when wet: The effect of a novel zwitterionic coating on silicone ureteral stents. Urolithiasis. 2026. doi:10.1007/s00240-026-02070-6