Under pressure: what we thought we knew about gravity irrigation in flexible ureteroscopy
Urolithiasis · 2025
DOI: 10.1007/s00240-025-01891-1 ↗Kurz gefasst
Fachartikel in der Zeitschrift Urolithiasis, 2025. Verfasst von Charlotte Slots und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, University Hospital Ghent, Corneel Heymanslaan 10, Ghent, Belgium. Charlotteslots@gmail.com. Hauptthemen laut MEDLINE: Ureteroscopy, Therapeutic Irrigation. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 10 Literaturangaben.
Abstract (Originalsprache)
This study revisits the physical principles of gravity-based irrigation in flexible ureteroscopy (fURS), assessing the effects of bag height, reference level, working channel occupancy, bag depletion, and drip chamber configuration on irrigation flow and pressure. An in vitro model was constructed with two 3-L saline bags connected to a Y-end system and a LithoVue™ fURS. Flow rates (mL/min) were measured at five heights (0-200 cm H₂O) using three reference levels: bag top, outflow, and drip chamber. Four working channel conditions were tested (free, 150 μm laser fiber, 200 μm fiber, and 1.9 Fr basket). Additional experiments examined bag volumes (3 L to near empty) and drip chamber content (air vs. fluid). Irrigation pressure at the ureteroscope tip was recorded via a water column. Flow increased proportionally with bag height and was lowest when referenced from the bag top. Channel occupancy markedly reduced flow, with no measurable flow at 0 cm when referenced from the top. Bag depletion and drip chamber air content had negligible influence, even in empty bags. Pressure recordings revealed a negative pressure in depleting bags, and the bag top correlated most closely with ureteroscope pressure. This confirms that gravity irrigation follows hydrostatic principles. Bag height and channel occupancy are the primary determinants of flow, whereas bag volume and drip chamber configuration exert minimal influence. The saline bag top best reflects irrigation pressure and should be the reference level. Negative pressures of 7-8 cm H₂O in depleting bags minimally affected flow but may impact intrarenal pressure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, University Hospital Ghent, Corneel Heymanslaan 10, Ghent, Belgium. Charlotteslots@gmail.com.
- Department of Urology, University Hospital Ghent, Corneel Heymanslaan 10, Ghent, Belgium.
- Department of Urology, AZ Delta Roeselare, Deltalaan 1, Roeselare, Belgium.
- Department of Urology, Hospital Clinic-University of Barcelona, Barcelona, Spain.
- Division of Urology, Department of Urology, Faculty of Medicine, Northern Ontario School of Medicine University, Benha University, Thunder Bay, ON, Egypt.
- Department of Urology, Sorbonne University, Tenon Hospital, AP-HP, Paris, France.
Zitieren
APA
Slots, C., Boeykens, M., Pauwaert, K., Sierra, A., Mylle, T., Noureldin, Y., Tailly, T., & Traxer, O. (2025). Under pressure: what we thought we knew about gravity irrigation in flexible ureteroscopy. Urolithiasis. https://doi.org/10.1007/s00240-025-01891-1
Vancouver
Slots C, Boeykens M, Pauwaert K, Sierra A, Mylle T, Noureldin Y, et al. Under pressure: what we thought we knew about gravity irrigation in flexible ureteroscopy. Urolithiasis. 2025. doi:10.1007/s00240-025-01891-1
BibTeX
@article{Slots2025,
author = {Slots, Charlotte and Boeykens, Matthias and Pauwaert, Kim and Sierra, Alba and Mylle, Toon and Noureldin, Yasser and Tailly, Thomas and Traxer, Olivier},
title = {Under pressure: what we thought we knew about gravity irrigation in flexible ureteroscopy},
journal = {Urolithiasis},
year = {2025},
doi = {10.1007/s00240-025-01891-1}
}