Feasibility, safety, and efficacy of a new percutaneous interspinous device: a retrospective multicenter study
Neuroradiology · 2024
DOI: 10.1007/s00234-024-03343-w ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Neuroradiology, 2024. Verfasst von Stefano Marcia und 8 weiteren Autor:innen. Beteiligte Einrichtung: UOC Radiologia SS, Trinità Hospital, 09121, Cagliari, Italy. Hauptthemen laut MEDLINE: Feasibility Studies, Spinal Stenosis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To evaluate safety and efficacy of the novel percutaneous interspinous device (PID) for the treatment of symptomatic degenerative lumbar spinal stenosis (DLSS) in 3 different centers. METHODS: From November 2016 to March 2020, 255 patients (male 125, mean age 71.2 years old range 49-91 years old) with neurogenic claudication, confirmed by electromyography, related to mono or bi-segmental lumbar central canal and/or foraminal stenosis were enrolled in the study. Magnetic resonance (MR) and/or computer tomography (CT), physical exam, and Visual Analogue Scale (VAS) and Zurich Claudication Questionnaire (ZCQ) were performed before and 6 months after the procedure. All treatments were performed under fluoroscopic guidance with local anesthesia and mild sedation. Technical success was defined as correct placement of the Lobster® (Demetrios Medical, Firenze, Italy) PID as demonstrated by computer tomography (CT) performed immediately after treatment; spinoplasty was performed in selected patients. RESULTS: PID placement was accomplished with a 99.6% success rate (257/258). The one device that was not implanted was due to a spinous process fracture. In 28 patients, more than 1 device was implanted in the same session (max 3 PIDs); 6 patients required a second implant in different session. A total of 172 prophylactic spinoplasties were performed (59.3%). No major complications occurred; 3 device misplacements were successfully treated with percutaneous retrieval and new device deployment. 99.6% of patients experienced clinical improvement. CONCLUSION: Lobster PID is an effective and safe minimally invasive decompression method for central canal and neural foraminal stenosis when patients are correctly selected.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- UOC Radiologia SS, Trinità Hospital, 09121, Cagliari, Italy.
- Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, MA, USA.
- UOC Neuroimmagini, Neuroradiologia Clinica E Funzionale Dipartimento Di Scienze Neurologiche E Motorie Azienda Ospedaliera Universitaria Senese, Siena, Italy.
- UOC Radiodiagnostica, Ospedali Riuniti Della Val Di Chiana, Montepulciano, Siena, Italy.
- Minimal Invasive Spine Dept of Neurosurgery, Istituto Oncologico del Mediterraneo IOM, Viagrande, Italy.
- Ingegneria Civile E Ambientale (DICEA), Università Di Firenze, Florence, Italy.
Zitieren
APA
Marcia, S., Hirsch, J. A., Bellini, M., Sadotti, G., Manfré, L., De Vivo, A. E., Piras, E., Zini, G., & Zini, C. (2024). Feasibility, safety, and efficacy of a new percutaneous interspinous device: a retrospective multicenter study. Neuroradiology. https://doi.org/10.1007/s00234-024-03343-w
Vancouver
Marcia S, Hirsch JA, Bellini M, Sadotti G, Manfré L, De Vivo AE, et al. Feasibility, safety, and efficacy of a new percutaneous interspinous device: a retrospective multicenter study. Neuroradiology. 2024. doi:10.1007/s00234-024-03343-w
BibTeX
@article{Marcia2024,
author = {Marcia, Stefano and Hirsch, Joshua Adam and Bellini, Matteo and Sadotti, Giulia and Manfré, Luigi and De Vivo, Aldo Eros and Piras, Emanuele and Zini, Giacomo and Zini, Chiara},
title = {Feasibility, safety, and efficacy of a new percutaneous interspinous device: a retrospective multicenter study},
journal = {Neuroradiology},
year = {2024},
doi = {10.1007/s00234-024-03343-w}
}
EUVIMED