Vergleichsstudie in der Zeitschrift Neuroradiology, 2024. Verfasst von Nana Fujita und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-Ku, Tokyo, 113-8655, Japan. Hauptthemen laut MEDLINE: Deep Learning, Temporal Bone, Tomography, X-Ray Computed, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The University of Tokyo.
Abstract (Originalsprache)
Purpose
We investigated whether the quality of high-resolution computed tomography (CT) images of the temporal bone improves with deep learning reconstruction (DLR) compared with hybrid iterative reconstruction (HIR).
Methods
This retrospective study enrolled 36 patients (15 men, 21 women; age, 53.9 ± 19.5 years) who had undergone high-resolution CT of the temporal bone. Axial and coronal images were reconstructed using DLR, HIR, and filtered back projection (FBP). In qualitative image analyses, two radiologists independently compared the DLR and HIR images with FBP in terms of depiction of structures, image noise, and overall quality, using a 5-point scale (5 = better than FBP, 1 = poorer than FBP) to evaluate image quality. The other two radiologists placed regions of interest on the tympanic cavity and measured the standard deviation of CT attenuation (i.e., quantitative image noise). Scores from the qualitative and quantitative analyses of the DLR and HIR images were compared using, respectively, the Wilcoxon signed-rank test and the paired t- test.
Results
Qualitative and quantitative image noise was significantly reduced in DLR images compared with HIR images (all comparisons, p ≤ 0.016). Depiction of the otic capsule, auditory ossicles, and tympanic membrane was significantly improved in DLR images compared with HIR images (both readers, p ≤ 0.003). Overall image quality was significantly superior in DLR images compared with HIR images (both readers, p < 0.001).
Conclusion
Compared with HIR, DLR provided significantly better-quality high-resolution CT images of the temporal bone.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-Ku, Tokyo, 113-8655, Japan.
- Department of Radiology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-Ku, Tokyo, 113-8655, Japan. koyasaka@gmail.com.
Förderung
- The University of Tokyo
Zitieren
APA
Fujita, N., Yasaka, K., Hatano, S., Sakamoto, N., Kurokawa, R., & Abe, O. (2024). Deep learning reconstruction for high-resolution computed tomography images of the temporal bone: comparison with hybrid iterative reconstruction. Neuroradiology. https://doi.org/10.1007/s00234-024-03330-1
Vancouver
Fujita N, Yasaka K, Hatano S, Sakamoto N, Kurokawa R, Abe O. Deep learning reconstruction for high-resolution computed tomography images of the temporal bone: comparison with hybrid iterative reconstruction. Neuroradiology. 2024. doi:10.1007/s00234-024-03330-1
BibTeX
@article{Fujita2024,
author = {Fujita, Nana and Yasaka, Koichiro and Hatano, Sosuke and Sakamoto, Naoya and Kurokawa, Ryo and Abe, Osamu},
title = {Deep learning reconstruction for high-resolution computed tomography images of the temporal bone: comparison with hybrid iterative reconstruction},
journal = {Neuroradiology},
year = {2024},
doi = {10.1007/s00234-024-03330-1}
}
EUVIMED