Interlaboratory comparison with a reference measurement procedure (RMP) for determining 24,25-dihydroxyvitamin D3 in human serum using liquid chromatography-tandem mass spectrometry (LC-MS/MS)
Analytical and Bioanalytical Chemistry · 28.04.2026
DOI: 10.1007/s00216-026-06480-8 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, online erschienen am 28.04.2026. Verfasst von Stephen A Wise und 14 weiteren Autor:innen. Beteiligte Einrichtung: IFC contractor in support of the Office of Dietary Supplements (ODS), National Institutes of Health (NIH), Bethesda, MD, 20817, USA. stephenawise@com…. Die Arbeit nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
An interlaboratory comparison study was conducted among five laboratories for determining 24,25-dihydroxyvitamin D 3 (24,25(OH) 2 D 3 ) in human serum using liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods. The Centers for Disease Control and Prevention (CDC), Imperial College Healthcare NHS Trust, University College Cork, University of Liège, and University of Washington analyzed 50 single-donor samples and two new Standard Reference Materials (SRMs®). The results from each laboratory were compared with target values assigned by the National Institute of Standards and Technology (NIST) using a reference measurement procedure (RMP) and evaluated using Ordinary Deming linear regression and Bland-Altman analysis. Three of the five laboratory methods provided results that were in good agreement with the NIST RMP results showing linear regression slopes ranging from 0.972 to 1.003 and Bland-Altman mean bias of −0.092 nmol/L, 0.025 nmol/L, and 0.035 nmol/L. Two laboratories demonstrated a significant positive bias with linear regression slopes of 1.158 and 1.214 and Bland-Altman mean bias of 0.162 nmol/L and 0.708 nmol/L. The CDC method is currently used to assign “information only” values for 24,25(OH) 2 D 3 in the quarterly distributions of the Vitamin D External Quality Assessment Scheme (DEQAS). Given the agreement (linear regression slope = 0.984, R 2 = 0.996 and mean bias of −0.092 nmol/L) between the CDC method and the NIST RMP observed in this study, the CDC-assigned 24,25(OH) 2 D 3 values in DEQAS may provide a more accurate reference than the current participant consensus mean values. Graphical Abstract
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- IFC contractor in support of the Office of Dietary Supplements (ODS), National Institutes of Health (NIH), Bethesda, MD, 20817, USA. stephenawise@comcast.net.
- Division of Laboratory Sciences, Centers for Disease Control and Prevention (CDC), Atlanta, GA, 30341, USA.
- Department of Clinical Chemistry, CIRM, CHU de Liège, University of Liège, 4000, Liège, Belgium.
- Cork Centre for Vitamin D and Nutrition Research, School of Food and Nutritional Sciences, University College Cork, Cork, T12 YT20, Ireland.
- Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, 98185, USA.
- Imperial College Healthcare NHS Trust, London, W6 8RF, UK.
Zitieren
APA
Wise, S. A., Mineva, E., Pfeiffer, C. M., Vesper, H., Cavalier, É., Peeters, S., Galvin, K., Cashman, K. D., Hoofnagle, A. N., Williams, E. L., Briggs, L. E., Souness, M. E., Kuszak, A. J., Hahm, G., & Camara, J. E. (2026). Interlaboratory comparison with a reference measurement procedure (RMP) for determining 24,25-dihydroxyvitamin D3 in human serum using liquid chromatography-tandem mass spectrometry (LC-MS/MS). Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-026-06480-8
Vancouver
Wise SA, Mineva E, Pfeiffer CM, Vesper H, Cavalier É, Peeters S, et al. Interlaboratory comparison with a reference measurement procedure (RMP) for determining 24,25-dihydroxyvitamin D3 in human serum using liquid chromatography-tandem mass spectrometry (LC-MS/MS). Analytical and Bioanalytical Chemistry. 2026. doi:10.1007/s00216-026-06480-8
BibTeX
@article{Wise2026,
author = {Wise, Stephen A and Mineva, Ekaterina and Pfeiffer, Christine M and Vesper, Hubert and Cavalier, Étienne and Peeters, Stephanie and Galvin, Karen and Cashman, Kevin D and Hoofnagle, Andrew N and Williams, Emma L and Briggs, Laura E and Souness, Megan E and Kuszak, Adam J and Hahm, Grace and Camara, Johanna E},
title = {Interlaboratory comparison with a reference measurement procedure (RMP) for determining 24,25-dihydroxyvitamin D3 in human serum using liquid chromatography-tandem mass spectrometry (LC-MS/MS)},
journal = {Analytical and Bioanalytical Chemistry},
year = {2026},
doi = {10.1007/s00216-026-06480-8}
}
EUVIMED