Demonstrating voxel-by-voxel (V × V) single-point calibration in liver tissue by IR-MALDESI quantitative MSI
Analytical and Bioanalytical Chemistry · 2025
DOI: 10.1007/s00216-025-06138-x ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, 2025. Verfasst von Emily R Bruce und 3 weiteren Autor:innen. Beteiligte Einrichtung: Biological Imaging Laboratory for Disease and Exposure Research, Department of Chemistry, North Carolina State University, Raleigh, NC, 27695, USA. Hauptthemen laut MEDLINE: Liver, Glutathione, Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIGMS NIH HHS, Division of Cancer Prevention, National Cancer Institute, National Institute of General Medical Sciences.
Abstract (Originalsprache)
Quantitative mass spectrometry imaging (qMSI) provides information regarding the colocalization, relative abundance, and concentration of a target analyte in a tissue without homogenization. Ionization sources, including IR-MALDESI, commonly utilize an on-tissue spatial calibration curve approach; however, this approach has several limitations including tedious sample preparation, and this approach does not account for local matrix effects. To compensate for these two limitations, we developed voxel-by-voxel (V × V) quantification to provide an internal standard calibration point for every voxel which requires a simple sample preparation and accounts for local matrix effects. In this work, we evaluate the performance of V × V quantification against the spatial calibration curve to assess the quantitative capacity of this newly developed method. Quantification of glutathione (GSH) on a per-voxel basis involves homogenously spraying a known amount of stable isotope-labeled glutathione (SIL-GSH) on a microscope slide. Next, we mount liver sections on top of the coated slides and image them using IR-MALDESI MSI. Statistical analysis demonstrated high precision for V × V quantification over a wide concentration range; however, the method’s accuracy is currently limited due to the sprayer’s configuration. Results support the feasibility of V × V quantification as evidenced by concentration heatmaps. Additionally, V × V quantification allows for parallel reaction monitoring (PRM) imaging which provides high specificity. Combined with relativity, straightforward sample preparation, and promising initial statistics, the V × V method offers significant advantages over spatial calibration curves. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Biological Imaging Laboratory for Disease and Exposure Research, Department of Chemistry, North Carolina State University, Raleigh, NC, 27695, USA.
- Department of Statistics, North Carolina State University, Raleigh, NC, 27695, USA.
- Biological Imaging Laboratory for Disease and Exposure Research, Department of Chemistry, North Carolina State University, Raleigh, NC, 27695, USA. dcmuddim@ncsu.edu.
Förderung
- NIGMS NIH HHS
- Division of Cancer Prevention, National Cancer Institute
- National Institute of General Medical Sciences
Zitieren
APA
Bruce, E. R., Kibbe, R. R., Opperman, L. J., & Muddiman, D. C. (2025). Demonstrating voxel-by-voxel (V × V) single-point calibration in liver tissue by IR-MALDESI quantitative MSI. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-025-06138-x
Vancouver
Bruce ER, Kibbe RR, Opperman LJ, Muddiman DC. Demonstrating voxel-by-voxel (V × V) single-point calibration in liver tissue by IR-MALDESI quantitative MSI. Analytical and Bioanalytical Chemistry. 2025. doi:10.1007/s00216-025-06138-x
BibTeX
@article{Bruce2025,
author = {Bruce, Emily R and Kibbe, Russell R and Opperman, Logan J and Muddiman, David C},
title = {Demonstrating voxel-by-voxel (V × V) single-point calibration in liver tissue by IR-MALDESI quantitative MSI},
journal = {Analytical and Bioanalytical Chemistry},
year = {2025},
doi = {10.1007/s00216-025-06138-x}
}
EUVIMED