EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Smartcard: an integrated approach for contaminant monitoring, from field to laboratory

Ariadni Geballa-Koukoula, Linda Willemsen, Erik Beij, Richard van Hoof, Alexander Elferink, Khalil Geballa-Koukoulas, Jeroen Peters, Marco H Blokland, Gert Ij Salentijn iD

Analytical and Bioanalytical Chemistry · 2025-01 · 417 (1) · 69-82

DOI: 10.1007/s00216-024-05626-w ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, online erschienen am 12.11.2024. Verfasst von Ariadni Geballa-Koukoula und 8 weiteren Autor:innen. Beteiligte Einrichtung: Wageningen Food Safety Research, Wageningen University and Research, P.O. Box 230, 6700 AE, Wageningen, The Netherlands. Hauptthemen laut MEDLINE: Food Contamination. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dutch Ministry of Agriculture, Nature, and Food Quality under their knowledge development program "Healthy and Safe Food Systems".

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Effective food safety monitoring requires a multi-step approach from farm to fork, involving different methods, ranging from convenient screening devices to sophisticated laboratory confirmatory testing. However, sample transportation to routine laboratories is time-consuming and expensive. Simplified on-site sampling followed by laboratory analysis offers a potential solution. Dried blood spot (DBS) cards ensure stability and ease of sample transportation and are used in clinical testing. However, the applicability of such an approach could be broader and include the storage of dried extract from more complex (solid) matrices. Therefore, a simplified approach is presented here, using DBS cards for on-site sampling and subsequent laboratory confirmation for food contaminants. To achieve this, an analytical tool (Smartcard) was designed using 3D-printing technology. As a proof of concept, the approach was applied to detect the pesticide fipronil, which is widely used in ornamental flower production to limit pests and on poultry farms. The Smartcard can securely store the sample extracts on a DBS card (dried extract spot (DExS) card), incorporate the lateral flow immunoassay (LFIA) and immediately provide an estimate of contamination levels. After simplified in-syringe extraction of the sample, the LFIA allows direct screening of fipronil (half maximum inhibitory concentration of 6.5 µg/l with calibration standards), and the same sample extract can be directly applied to the DExS card for storage and transport to the laboratory, where analyte re-extraction and instrumental analysis is performed using ultra high-pressure liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) detecting fipronil down to 0.8 µg/kg. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Food ContaminationPyrazolesDried Blood Spot TestingAnimalsTandem Mass SpectrometryLimit of DetectionImmunoassayInsecticides

Stichwörter der Autor:innen

Dried blood spot (DBS) cardFipronilLateral flow immunoassay (LFIA)On-site application

Substanzen

fipronilPyrazolesInsecticides

Autor:innen und Institutionen

  • Wageningen Food Safety Research, Wageningen University and Research, P.O. Box 230, 6700 AE, Wageningen, The Netherlands.
  • Laboratory of General and Agricultural Microbiology, Department of Crop Science, Agricultural University of Athens, Iera Odos 75, 118 55, Athens, Greece.
  • Wageningen Food Safety Research, Wageningen University and Research, P.O. Box 230, 6700 AE, Wageningen, The Netherlands. gert.salentijn@wur.nl.
  • Laboratory of Organic Chemistry, Wageningen University, Stippeneng 4, 6708 WE, Wageningen, The Netherlands. gert.salentijn@wur.nl.

ORCID-Profile: Gert Ij Salentijn

Förderung

  • Dutch Ministry of Agriculture, Nature, and Food Quality under their knowledge development program "Healthy and Safe Food Systems"

Zitieren

APA

Geballa-Koukoula, A., Willemsen, L., Beij, E., van Hoof, R., Elferink, A., Geballa-Koukoulas, K., Peters, J., Blokland, M. H., & Salentijn, G. I. (2025). Smartcard: an integrated approach for contaminant monitoring, from field to laboratory. Analytical and Bioanalytical Chemistry, 417(1), 69-82. https://doi.org/10.1007/s00216-024-05626-w

Vancouver

Geballa-Koukoula A, Willemsen L, Beij E, van Hoof R, Elferink A, Geballa-Koukoulas K, et al. Smartcard: an integrated approach for contaminant monitoring, from field to laboratory. Analytical and Bioanalytical Chemistry. 2025;417(1):69-82. doi:10.1007/s00216-024-05626-w

BibTeX
@article{GeballaKoukoula2025,
  author = {Geballa-Koukoula, Ariadni and Willemsen, Linda and Beij, Erik and van Hoof, Richard and Elferink, Alexander and Geballa-Koukoulas, Khalil and Peters, Jeroen and Blokland, Marco H and Salentijn, Gert Ij},
  title = {Smartcard: an integrated approach for contaminant monitoring, from field to laboratory},
  journal = {Analytical and Bioanalytical Chemistry},
  year = {2025},
  volume = {417},
  number = {1},
  pages = {69--82},
  doi = {10.1007/s00216-024-05626-w}
}
Kostenlos starten →