The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study
Clinical Neuroradiology · 2025
DOI: 10.1007/s00062-024-01446-8 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Clinical Neuroradiology, 2025. Verfasst von Roland Schwab und 14 weiteren Autor:innen. Beteiligte Einrichtung: University Clinic for Neuroradiology, Medical Faculty and University Hospital, Otto-von-Guericke-University Magdeburg, Magdeburg, Germany. Schwab.Rol…. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Embolization, Therapeutic, Aneurysm, Ruptured. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Otto-von-Guericke-Universität Magdeburg.
Abstract (Originalsprache)
Backround The use of flow diverting stents in the treatment of intracranial aneurysms is associated with a risk of neurological morbidity due to their thrombogenicity. To reduce this risk different surface modifications have been developed. The Derivo 2 Embolization Device (Acandis, Pforzheim, Germany) has proven to be a safe and effective flow diverter. To overcome the risk of thrombo-embolism, the device was modified by adding an anti-thrombogenic fibrin-heparin coating. We aimed to assess the safety and effectiveness of the Derivo 2 heal Embolization Device.
Methods
Retrospective multicenter data from nine German neurovascular centers between February 2022 until December 2023 were used. Patients treated with the Derivo 2 heal Embolization Device for unruptured or ruptured intracranial aneurysms were included. Peri- and postprocedural adverse events, clinical outcomes, and angiographic follow-up results were evaluated.
Results
84 patients (73.8% female; mean age 58.7 years) with 89 aneurysms (mean size 9.8 mm) were included. 87.6% were located in the anterior circulation. Most of them were sidewall aneurysms (88.8%). 96 flow diverters were used. 99.0% were successfully implanted. An in-stent balloon angioplasty was performed in 6.0% of the cases. An additional coiling was performed in 28.6%. Technical difficulties were present in 12.0% of the cases. Thrombotic events occurred in 4.8% with no neurological sequelae. Mortality and morbidity were 0 and 1.2% respectively. Adequate aneurysm occlusion was achieved in 80.7% with a mean follow-up time of 6.6 months.
Conclusion
The Derivo 2 heal Embolization Device showed a satisfying aneurysm occlusion and safety with a low rate of neurological morbidity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University Clinic for Neuroradiology, Medical Faculty and University Hospital, Otto-von-Guericke-University Magdeburg, Magdeburg, Germany. Schwab.Roland@outlook.de.
- Research Campus STIMULATE, Otto-von-Guericke University Magdeburg, Magdeburg, Germany. Schwab.Roland@outlook.de.
- Department of Radiology and Neuroradiology, Faculty of Medicine and University Hospital, University of Cologne, Cologne, Germany.
- Department of Diagnostic and Interventional Radiology, Medical Faculty, University Düsseldorf, Düsseldorf, Germany.
- Department of Radiology and Neuroradiology, Klinikum Vest, Recklinghausen, Germany.
- Department of Diagnostic and Interventional Neuroradiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Förderung
- Otto-von-Guericke-Universität Magdeburg
Zitieren
APA
Schwab, R., Kabbasch, C., Goertz, L., Kaschner, M., Weiss, D., Loehr, C., Wensing, H., Bester, M., Simgen, A., Kemmling, A., Wendl, C., Fuchs, E., Thormann, M., Behme, D., & Nordmeyer, H. (2025). The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study. Clinical Neuroradiology. https://doi.org/10.1007/s00062-024-01446-8
Vancouver
Schwab R, Kabbasch C, Goertz L, Kaschner M, Weiss D, Loehr C, et al. The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study. Clinical Neuroradiology. 2025. doi:10.1007/s00062-024-01446-8
BibTeX
@article{Schwab2025,
author = {Schwab, Roland and Kabbasch, Christoph and Goertz, Lukas and Kaschner, Marius and Weiss, Daniel and Loehr, Christian and Wensing, Hauke and Bester, Maxim and Simgen, Andreas and Kemmling, André and Wendl, Christina and Fuchs, Erelle and Thormann, Maximilian and Behme, Daniel and Nordmeyer, Hannes},
title = {The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study},
journal = {Clinical Neuroradiology},
year = {2025},
doi = {10.1007/s00062-024-01446-8}
}