Validation of a quantitative immunochromatographic field test for measuring IgG in calf blood
Veterinary Record · 19.11.2026
DOI: 10.1002/vetr.71373 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Record, online erschienen am 19.11.2026. Verfasst von Nick Britten und 7 weiteren Autor:innen. Die Arbeit nennt 38 Literaturangaben.
Abstract (Originalsprache)
Background
Identifying failed transfer of passive immunity (FTPI) through the IgG status of calves is important for health and production. Rapid and accurate assessment of IgG status is desirable, but currently available methods have limitations. This study compares the performance of a novel immunochromatographic field test for IgG with a laboratory analysis for calves aged 1‒28 days using heparinised whole blood.
Methods
Excess blood collected from calves aged 1‒28 days on dairy farms in south‐west England was used to compare the novel immunochromatographic field test with a commercially available ELISA. Bland‒Altman analysis, Pearson's correlation coefficient and receiver operating characteristic curves were used to evaluate agreement between the two tests.
Results
Pearson's correlation between the two tests was 0.78, showing a strong relationship. The ELISA returned slightly higher IgG concentrations than the field test, but the relationship was proportionally consistent across the range of values. The field test had a sensitivity and specificity of 86.7% and 95.6%, respectively, for the detection of FTPI at a threshold of 9.3 mg/mL. Test performance was consistent across the range of calf ages.
Limitations
There was no systematic selection of calves. Sex, breed, farm and other variables were not available for analysis. Jugular samples were used rather than the recommended nose‐prick samples.
Conclusion
The field test showed consistently strong agreement with the ELISA test across ages of calves and IgG values. Sensitivity and specificity in detecting FTPI were very good.
Zitieren
APA
Britten, N., Fisher, M. A., Fenn, C. A., Middleton, B., Hayton, A., Payne, J., Hudson, C., & Hayton, R. (2026). Validation of a quantitative immunochromatographic field test for measuring IgG in calf blood. Veterinary Record. https://doi.org/10.1002/vetr.71373
Vancouver
Britten N, Fisher MA, Fenn CA, Middleton B, Hayton A, Payne J, et al. Validation of a quantitative immunochromatographic field test for measuring IgG in calf blood. Veterinary Record. 2026. doi:10.1002/vetr.71373
BibTeX
@article{Britten2026,
author = {Britten, Nick and Fisher, Maggie A. and Fenn, Caroline A. and Middleton, Brian and Hayton, Alastair and Payne, Joseph and Hudson, Chris and Hayton, Rachel},
title = {Validation of a quantitative immunochromatographic field test for measuring IgG in calf blood},
journal = {Veterinary Record},
year = {2026},
doi = {10.1002/vetr.71373}
}
EUVIMED