Evaluating the accuracy of a six‐lead smartphone‐based electrocardiographic device compared with standard electrocardiography in brachymorphic dogs
Veterinary Record · 07.10.2023 · 193 (7)
DOI: 10.1002/vetr.2879 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Veterinary Record, online erschienen am 01.05.2023. Verfasst von Giovanni Romito und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Medical Sciences, Alma Mater Studiorum-University of Bologna, Ozzano dell'Emilia, Italy. Hauptthemen laut MEDLINE: Smartphone, Dog Diseases. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
No previous study has evaluated the accuracy of a six‐lead smartphone‐based electrocardiographic (s‐ECG) device in dogs.
Methods
This was a prospective multicentre study. Patients referred for cardiologic consultation were selected. In each patient, a 30‐second electrocardiogram was simultaneously acquired with a novel six‐lead s‐ECG and a standard six‐lead ECG machine (st‐ECG). A board‐certified cardiologist evaluated each recording. Nineteen ECG variables, including heart rate and rhythm, as well as quantitative and qualitative features of waves, segments and intervals, were analysed. Agreement between s‐ECG and st‐ECG was evaluated using Cohen's kappa coefficient and the Bland–Altman test.
Results
Seventy‐five dogs were enrolled, and 140 ECG tracings were analysed. There was perfect agreement between the two methodologies for heart rate and rhythm classification, both in dogs with sinus rhythm and those with pathological rhythms. Although some disagreement was found when comparing measurements of quantitative variables obtained with the s‐ECG and the st‐ECG, none of the differences was of clinical relevance.
Limitations
The sample size was limited, and the interobserver variability was not analysed.
Conclusion
The six‐lead s‐ECG studied herein is comparable to the st‐ECG for heart rate and rhythm assessment, and seems clinically acceptable for the interpretation of waves, segments and intervals in dogs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Veterinary Medical Sciences, Alma Mater Studiorum-University of Bologna, Ozzano dell'Emilia, Italy.
- Freelance Veterinary Cardiologist, Bologna, Italy.
- Department of Veterinary Sciences, University of Parma, Parma, Italy.
ORCID-Profile: Giovanni Romito, Maria Chiara Sabetti, Michela Ablondi, Mario Cipone
Zitieren
APA
Romito, G., Castagna, P., Sabetti, M. C., Ablondi, M., & Cipone, M. (2023). Evaluating the accuracy of a six‐lead smartphone‐based electrocardiographic device compared with standard electrocardiography in brachymorphic dogs. Veterinary Record, 193(7). https://doi.org/10.1002/vetr.2879
Vancouver
Romito G, Castagna P, Sabetti MC, Ablondi M, Cipone M. Evaluating the accuracy of a six‐lead smartphone‐based electrocardiographic device compared with standard electrocardiography in brachymorphic dogs. Veterinary Record. 2023;193(7). doi:10.1002/vetr.2879
BibTeX
@article{Romito2023,
author = {Romito, Giovanni and Castagna, Prisca and Sabetti, Maria Chiara and Ablondi, Michela and Cipone, Mario},
title = {Evaluating the accuracy of a six‐lead smartphone‐based electrocardiographic device compared with standard electrocardiography in brachymorphic dogs},
journal = {Veterinary Record},
year = {2023},
volume = {193},
number = {7},
doi = {10.1002/vetr.2879}
}
EUVIMED