Accuracy of portable ultrasound machines for obstetric biometry
Ultrasound in Obstetrics & Gynecology · 2024
DOI: 10.1002/uog.27541 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Ultrasound in Obstetrics & Gynecology, 2024. Verfasst von K J Rittenhouse und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Obstetrics and Gynecology, University of North Carolina School of Medicine, Chapel Hill, NC, USA. Hauptthemen laut MEDLINE: Ultrasonography, Prenatal, Gestational Age, Biometry. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: FIC NIH HHS, NIAID NIH HHS, NICHD NIH HHS.
Abstract (Originalsprache)
Objective
To evaluate the accuracy of two portable ultrasound machines (PUM) in assessing fetal biometry and estimated gestational age (EGA).
Methods
This was a secondary analysis of data from the Fetal Age Machine Learning Initiative, an observational study of pregnant women in the USA and Zambia. Each participant underwent ultrasound assessment by an experienced sonographer using both a high‐specification ultrasound machine (HSUM) and a PUM (Butterfly iQ or Clarius C3) to measure fetal biometry and calculate EGA at each visit. By comparing paired PUM and HSUM scans, we estimated agreement between individual biometry measurements and aggregate gestational age estimates by reporting mean difference, intraclass correlation coefficient (ICC) and Bland–Altman plots, adjusting for trend.
Results
Between April and December 2021, 818 participants contributed 1386 paired PUM–HSUM ultrasound investigations, of which 991 PUM scans were obtained using the Butterfly iQ device and 395 using the Clarius C3 device. Gestational age at scan ranged from 7 to 38 weeks. Compared with HSUM, the Butterfly iQ PUM had a mean difference of −0.20 (95% CI, −0.60 to 0.20) days in the first trimester and −0.68 (95% CI, −0.93 to −0.44) days in the second/third trimesters. Compared with HSUM, the Clarius C3 PUM had a mean difference of −0.47 (95% CI, −1.11 to 0.18) days in the first trimester and −1.67 (95% CI, −2.10 to −1.25) days in the second/third trimesters. ICCs were 0.989 or greater throughout. Increasing gestational age was associated with increasing error and absolute error in EGA and fetal biometry. Both PUM devices demonstrated a modest trend toward underestimation of EGA with advancing gestational age in second/third‐trimester scans, compared with HSUM.
Conclusion
The Butterfly iQ and Clarius C3 PUM devices were highly accurate in performing fetal biometry in a diverse population from the USA and Zambia. © 2023 International Society of Ultrasound in Obstetrics and Gynecology.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Obstetrics and Gynecology, University of North Carolina School of Medicine, Chapel Hill, NC, USA.
- Department of Obstetrics and Gynaecology, University of Zambia School of Medicine, Lusaka, Zambia.
- Department of Biostatistics, University of North Carolina Gillings School of Global Public Health, Chapel Hill, NC, USA.
- UNC Global Projects - Zambia LLC, Lusaka, Zambia.
- Department of Psychiatry, University of North Carolina School of Medicine, Chapel Hill, NC, USA.
- Department of Epidemiology, University of North Carolina Gillings School of Global Public Health, Chapel Hill, NC, USA.
ORCID-Profile: K J Rittenhouse
Förderung
- FIC NIH HHS
- NIAID NIH HHS
- NICHD NIH HHS
- Bill and Melinda Gates Foundation
- NIH HHS
- National Institutes of Health
Zitieren
APA
Rittenhouse, K. J., Vwalika, B., Sebastião, Y., Pokaprakarn, T., Sindano, N., Shah, H., Stringer, E. M., Kasaro, M. P., Cole, S. R., Stringer, J. S. A., & Price, J. T. (2024). Accuracy of portable ultrasound machines for obstetric biometry. Ultrasound in Obstetrics & Gynecology. https://doi.org/10.1002/uog.27541
Vancouver
Rittenhouse KJ, Vwalika B, Sebastião Y, Pokaprakarn T, Sindano N, Shah H, et al. Accuracy of portable ultrasound machines for obstetric biometry. Ultrasound in Obstetrics & Gynecology. 2024. doi:10.1002/uog.27541
BibTeX
@article{Rittenhouse2024,
author = {Rittenhouse, K J and Vwalika, B and Sebastião, Y and Pokaprakarn, T and Sindano, N and Shah, H and Stringer, E M and Kasaro, M P and Cole, S R and Stringer, J S A and Price, J T},
title = {Accuracy of portable ultrasound machines for obstetric biometry},
journal = {Ultrasound in Obstetrics & Gynecology},
year = {2024},
doi = {10.1002/uog.27541}
}
EUVIMED