EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Electrochemical Biosensor Based on Hairy Core–Shell Particles: Effect of Core Conductivity

Pavel Milkin, Anila Antony, Hongtao Cai, Ceyda Topal, Antonia Debevc, Anne Linhardt, Alla Synytska iD, Leonid Ionov iD

Small Science · 2026-04 · 6 (4)

DOI: 10.1002/smsc.202500550 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Small Science, online erschienen am 07.04.2026. Verfasst von Pavel Milkin und 7 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Engineering Sciences University of Bayreuth Bayreuth Germany. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, the effect of the core material in hairy core–shell carriers with grafted poly(2‐(dimethylamino)ethyl methacrylate) (PDMAEMA) polymer brushes containing immobilized laccase from Trametes versicolor ( TvL ) is investigated. Conductive silver, silver‐Janus, carbon nanotubes, carbon black, and insulating silica particles were chosen as core materials. These carriers are easy to handle and store, enabling reproducible sensor fabrication with a well‐defined and high number of immobilized enzymes. Hydroquinone (HQ) detection in an aqueous system was chosen as a model for biosensor characterization. All systems demonstrated high catalytic efficiency, significantly suppressing the oxidation of HQ at the electrode surface and thereby providing selectivity. The sensitivity of the prepared sensors did not differ significantly among the carriers. However, the limit of detection was highly dependent on the overall conductivity of the carrier and its active surface area. Our main observation is that using highly conductive, high‐surface‐area carriers does not necessarily enhance sensor performance and can in fact worsen the detection limit due to the dominance of capacitive currents over faradaic ones.

Schlagwörter

Stichwörter der Autor:innen

bioelectrocatalysiselectrochemical biosensorhairy core–shell particleslaccasepolymer brush

Autor:innen und Institutionen

  • Faculty of Engineering Sciences University of Bayreuth Bayreuth Germany.
  • Functional Polymer Interfaces Group Bayerisches Polymerinstitut (BPI) University of Bayreuth Bayreuth Germany.

ORCID-Profile: Alla Synytska, Leonid Ionov

Förderung

  • Deutsche Forschungsgemeinschaft

Zitieren

APA

Milkin, P., Antony, A., Cai, H., Topal, C., Debevc, A., Linhardt, A., Synytska, A., & Ionov, L. (2026). Electrochemical Biosensor Based on Hairy Core–Shell Particles: Effect of Core Conductivity. Small Science, 6(4). https://doi.org/10.1002/smsc.202500550

Vancouver

Milkin P, Antony A, Cai H, Topal C, Debevc A, Linhardt A, et al. Electrochemical Biosensor Based on Hairy Core–Shell Particles: Effect of Core Conductivity. Small Science. 2026;6(4). doi:10.1002/smsc.202500550

BibTeX
@article{Milkin2026,
  author = {Milkin, Pavel and Antony, Anila and Cai, Hongtao and Topal, Ceyda and Debevc, Antonia and Linhardt, Anne and Synytska, Alla and Ionov, Leonid},
  title = {Electrochemical Biosensor Based on Hairy Core–Shell Particles: Effect of Core Conductivity},
  journal = {Small Science},
  year = {2026},
  volume = {6},
  number = {4},
  doi = {10.1002/smsc.202500550}
}
Kostenlos starten →