Machine Learning‐Driven Capillary Microfluidic Design Automation for Programmable Gradient Generation and Antimicrobial Testing

Mahmood Khalghollah, Azam Zare, Sorosh Abdollahi iD, Behrouz H Far, Amir Sanati Nezhad iD

Small · 2026-07 · 22 (37)

DOI: 10.1002/smll.73821 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Small, online erschienen am 17.05.2026. Verfasst von Mahmood Khalghollah und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Software Engineering, University of Calgary, Calgary, Alberta, Canada. Hauptthemen laut MEDLINE: Machine Learning, Microfluidics, Anti-Infective Agents. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Canada Research Chairs Program, Natural Sciences and Engineering Research Council of Canada, NSERC CREATE Training Program in Wearable Technology Research and Collaboration.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rapid, portable, and automated determination of minimum inhibitory concentration (MIC) is critical for antimicrobial susceptibility testing in decentralized settings. We present TCG‐CMDA (Tree‐Shaped Concentration Gradient generator enabled by Capillary Microfluidic Design Automation), a machine learning–guided platform for the automated design of capillary microfluidic chips that passively generate programmable concentration gradients. A key innovation of TCG‐CMDA is its ability to achieve fully synchronized, capillary‐driven flow of two preloaded agents, enabling precise and reproducible gradient formation without external pumps or valves. By integrating computational fluid dynamics (CFD) simulations, a neural network surrogate model, and quasi‐Newton optimization, the platform customizes tree‐shaped geometries to match target mixing profiles. We validate TCG‐CMDA through dye flow visualization and MIC testing of Escherichia coli against gentamicin, showing high concordance with conventional antibiotic susceptibility testing. The resulting chip supports a streamlined workflow consisting of passive loading, autonomous gradient generation, and incubation (∼2 h), followed by direct endpoint assessment, making it well suited for decentralized point‐of‐care applications. TCG‐CMDA establishes a closed‐loop, simulation‐guided design paradigm for capillary microfluidic systems. Although demonstrated here for MIC‐oriented antimicrobial testing, the framework is generalizable and can be extended to other gradient‐dependent microfluidic applications through geometry reconfiguration and fluid‐specific optimization.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Machine LearningMicrofluidicsAnti-Infective AgentsMicrobial Sensitivity TestsAutomationEscherichia coliHydrodynamicsEquipment Design

Stichwörter der Autor:innen

capillary microfluidicdesign automationmachine learningminimum inhibitory concentration testingneural networktree‐shaped concentration gradient generator

Substanzen

Anti-Infective Agents

Autor:innen und Institutionen

  • Department of Electrical and Software Engineering, University of Calgary, Calgary, Alberta, Canada.
  • BioMEMS and Bioinspired Microfluidic Laboratory, Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Alberta, Canada.
  • Department of Biomedical Engineering, Schulich School of Engineering, University of Calgary, Calgary, Alberta, Canada.
  • Department of Mechanical and Manufacturing Engineering, University of Calgary, Calgary, Alberta, Canada.

ORCID-Profile: Sorosh Abdollahi, Amir Sanati Nezhad

Förderung

  • Canada Research Chairs Program
  • Natural Sciences and Engineering Research Council of Canada
  • NSERC CREATE Training Program in Wearable Technology Research and Collaboration

Zitieren

APA

Khalghollah, M., Zare, A., Abdollahi, S., Far, B. H., & Sanati Nezhad, A. (2026). Machine Learning‐Driven Capillary Microfluidic Design Automation for Programmable Gradient Generation and Antimicrobial Testing. Small, 22(37). https://doi.org/10.1002/smll.73821

Vancouver

Khalghollah M, Zare A, Abdollahi S, Far BH, Sanati Nezhad A. Machine Learning‐Driven Capillary Microfluidic Design Automation for Programmable Gradient Generation and Antimicrobial Testing. Small. 2026;22(37). doi:10.1002/smll.73821

BibTeX
@article{Khalghollah2026,
  author = {Khalghollah, Mahmood and Zare, Azam and Abdollahi, Sorosh and Far, Behrouz H and Sanati Nezhad, Amir},
  title = {Machine Learning‐Driven Capillary Microfluidic Design Automation for Programmable Gradient Generation and Antimicrobial Testing},
  journal = {Small},
  year = {2026},
  volume = {22},
  number = {37},
  doi = {10.1002/smll.73821}
}