A Semi‐automatic Puncture System for Liver Rupture Haemostasis: Ex Vivo and In Vivo Study
The International Journal of Medical Robotics and Computer Assisted Surgery · 2026-12 · 22 (6)
DOI: 10.1002/rcs.70248 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, online erschienen am 06.11.2026. Verfasst von Pak Kin Wong und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Macau, Macau, China. Hauptthemen laut MEDLINE: Liver, Punctures, Robotic Surgical Procedures. Die Arbeit nennt 22 Literaturangaben. Gefördert von: National Natural Science Foundation of China, The China Postdoctoral Science Foundation, Guangdong Basic and Applied Basic Research Foundation.
Abstract (Originalsprache)
Background
Liver rupture caused by blunt trauma can result in severe haemorrhage and death. Percutaneous puncture embolisation is a potential minimally invasive approach for haemostasis, but manual puncture is limited by difficult bleeding‐site localization and hand–eye coordination.
Methods
We developed a semi‐automatic puncture system integrating a robotic arm and a compact rack‐pinion driven end‐effector. Positioning and posture accuracy were evaluated, followed by ex vivo and in vivo porcine liver embolisation experiments.
Results
The system can achieve mean positioning and posture accuracies of 0.36 mm and 1.03° under load‐free conditions, and 1.83 mm and 2.67° in the soft‐tissue phantom, respectively. Embolic material was delivered to the target region, and cessation of visible bleeding was observed within approximately 20 s in the ex vivo and in vivo experiments.
Conclusions
The results support the preliminary feasibility of this approach for percutaneous puncture embolisation of complex hepatic lacerations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Macau, Macau, China.
- Department of Electromechanical Engineering, University of Macau, Macau, China.
- Tianjin Key Laboratory of Composite and Functional Materials, School of Materials Science and Engineering, Tianjin University, Tianjin, China.
- Harbin Institute of Technology, Shenzhen, China.
- Southern University of Science and Technology, Shenzhen, China.
- Shenzhen Key Laboratory of Minimally Invasive Surgical Robotics and System, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
ORCID-Profile: Pak Kin Wong, Jinhang Li, Peixuan Ge, Ziqi Zhou, Liang Yao, Baoliang Zhao, Peng Zhang
Förderung
- National Natural Science Foundation of China
- The China Postdoctoral Science Foundation
- Guangdong Basic and Applied Basic Research Foundation
- Shenzhen Science and Technology Program
Zitieren
APA
Wong, P. K., Li, J., Ge, P., Ji, Y., Zhou, Z., Zhang, T., Yao, L., Zhao, B., Zhang, P., & Hu, Y. (2026). A Semi‐automatic Puncture System for Liver Rupture Haemostasis: Ex Vivo and In Vivo Study. The International Journal of Medical Robotics and Computer Assisted Surgery, 22(6). https://doi.org/10.1002/rcs.70248
Vancouver
Wong PK, Li J, Ge P, Ji Y, Zhou Z, Zhang T, et al. A Semi‐automatic Puncture System for Liver Rupture Haemostasis: Ex Vivo and In Vivo Study. The International Journal of Medical Robotics and Computer Assisted Surgery. 2026;22(6). doi:10.1002/rcs.70248
BibTeX
@article{Wong2026,
author = {Wong, Pak Kin and Li, Jinhang and Ge, Peixuan and Ji, Yunpeng and Zhou, Ziqi and Zhang, Tao and Yao, Liang and Zhao, Baoliang and Zhang, Peng and Hu, Ying},
title = {A Semi‐automatic Puncture System for Liver Rupture Haemostasis: Ex Vivo and In Vivo Study},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2026},
volume = {22},
number = {6},
doi = {10.1002/rcs.70248}
}
EUVIMED