Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, online erschienen am 29.11.2025. Verfasst von Alexander D Orsi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Corin, Clinical Innovation, Raynham, Massachusetts, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

This study examines the agreement and error between manual gap assessment, and assessment with a robotic soft‐tissue tensioning device in total knee arthroplasty across multiple surgeons.

Methods

517 consecutive robotic‐assisted TKAs were retrospectively reviewed across six surgeons following ethics approval. Robotic and manual mediolateral (ML) balance and medial and lateral gaps were compared using Intraclass Correlation Coefficient (ICC). ΔML balance (manual‐robotic) was compared between right and left knees.

Results

Manual and robotic ML balance showed poor agreement (ICC = 0.368). Surgeons had low to moderate correlations (ICC range: 0.152–0.633) and significant differences throughout flexion. Four measured greater medial gaps manually and two measured greater lateral gaps compared to robotics. All showed significant ΔML balance differences between the leg sides ( p < 0.05), with surgeon‐specific variations in gap patterns.

Conclusions

Manual ML balance correlated poorly and differed significantly from robotic assessment, with surgeon‐specific variation. Surgeons should appreciate the variability in subjective versus objective gap measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeRobotic Surgical ProceduresHumansRetrospective StudiesFemaleMaleAgedMiddle AgedKnee JointRange of Motion, Articular

Stichwörter der Autor:innen

computer assisted surgeryligament balancingrobotic assisted surgerytotal knee arthroplasty

Autor:innen und Institutionen

  • Corin, Clinical Innovation, Raynham, Massachusetts, USA.
  • Department of Orthopaedics, Nepean Hospital, Penrith, New South Wales, Australia.
  • Orthopaedic Associates of Reading, Wyomissing, Pennsylvania, USA.
  • Sierra Pacific Orthopedics, Fresno, California, USA.
  • Gundersen Health System, Viroqua, Wisconsin, USA.
  • Orthopaedics New England, Middlebury, Connecticut, USA.

ORCID-Profile: Alexander D Orsi, Christopher Plaskos

Zitieren

APA

Orsi, A. D., Coffey, S., Slotkin, E., Gill, P., Lehman, A., Lawrence, J., Plaskos, C., & Keggi, J. (2025). Soft Tissue Balance Profiles Differ Between Manual and Robotically Assessed Gaps in Total Knee Arthroplasty. The International Journal of Medical Robotics and Computer Assisted Surgery, 21(6). https://doi.org/10.1002/rcs.70124

Vancouver

Orsi AD, Coffey S, Slotkin E, Gill P, Lehman A, Lawrence J, et al. Soft Tissue Balance Profiles Differ Between Manual and Robotically Assessed Gaps in Total Knee Arthroplasty. The International Journal of Medical Robotics and Computer Assisted Surgery. 2025;21(6). doi:10.1002/rcs.70124

BibTeX
@article{Orsi2025,
  author = {Orsi, Alexander D and Coffey, Simon and Slotkin, Eric and Gill, Paramjeet and Lehman, Andrew and Lawrence, Jeffrey and Plaskos, Christopher and Keggi, John},
  title = {Soft Tissue Balance Profiles Differ Between Manual and Robotically Assessed Gaps in Total Knee Arthroplasty},
  journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
  year = {2025},
  volume = {21},
  number = {6},
  doi = {10.1002/rcs.70124}
}
Kostenlos starten →