A haptic guidance system for simulated catheter navigation with different kinaesthetic feedback profiles
The International Journal of Medical Robotics and Computer Assisted Surgery · 2024
DOI: 10.1002/rcs.2638 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2024. Verfasst von Taha Abbasi-Hashemi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical, Computer, and Biomedical Engineering, Toronto Metropolitan University, Toronto, Ontario, Canada. Hauptthemen laut MEDLINE: Surgery, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 58 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council opf Canada.
Abstract (Originalsprache)
Background
This paper proposes a haptic guidance system to improve catheter navigation within a simulated environment.
Methods
Three force profiles were constructed to evaluate the system: collision prevention; centreline navigation; and a novel force profile of reinforcement learning (RL). All force profiles were evaluated from the left common iliac to the right atrium.
Results
Our findings show that providing haptic feedback improved surgical safety compared to visual‐only feedback. If staying inside the vasculature is the priority, RL provides the safest option. It is also shown that the performance of each force profile varies in different anatomical regions.
Conclusion
The implications of these findings are significant, as they hold the potential to improve how and when haptic feedback is applied for cardiovascular intervention.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical, Computer, and Biomedical Engineering, Toronto Metropolitan University, Toronto, Ontario, Canada.
- Department of Mechanical, Industrial and Mechatronics Engineering, Toronto Metropolitan University, Toronto, Ontario, Canada.
- Cardiology, St. Michael's Hospital, Toronto, Ontario, Canada.
ORCID-Profile: Taha Abbasi-Hashemi, Farrokh Janabi-Sharifi, Asim N Cheema, Kourosh Zareinia
Förderung
- Natural Sciences and Engineering Research Council opf Canada
Zitieren
APA
Abbasi-Hashemi, T., Janabi-Sharifi, F., Cheema, A. N., & Zareinia, K. (2024). A haptic guidance system for simulated catheter navigation with different kinaesthetic feedback profiles. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.2638
Vancouver
Abbasi-Hashemi T, Janabi-Sharifi F, Cheema AN, Zareinia K. A haptic guidance system for simulated catheter navigation with different kinaesthetic feedback profiles. The International Journal of Medical Robotics and Computer Assisted Surgery. 2024. doi:10.1002/rcs.2638
BibTeX
@article{AbbasiHashemi2024,
author = {Abbasi-Hashemi, Taha and Janabi-Sharifi, Farrokh and Cheema, Asim N and Zareinia, Kourosh},
title = {A haptic guidance system for simulated catheter navigation with different kinaesthetic feedback profiles},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2024},
doi = {10.1002/rcs.2638}
}
EUVIMED